Laboratory Medicine ›› 2015, Vol. 30 ›› Issue (4): 324-326.DOI: 10.3969/j.issn.1673-8640.2015.04.007

Previous Articles     Next Articles

Evaluation on the anti-lipid turbidity capability of D-dimer by immunonephelometry using EP7-A2

NIE Bin, TAO Tianlu, HUANG Tao, XIAO Huayong.   

  1. Department of Clinical Laboratory, the Second People's Hospital of Yibin, Sichuan Yibin 644000, China
  • Received:2014-11-07 Online:2015-04-30 Published:2015-05-16

Abstract:

Objective To evaluate the anti-lipid turbidity capability of D-dimer by immunonephelometry using EP7-A2. Methods According to EP7-A2, paired difference testing and dosage effect testing were performed. Results The chylus with 1 200 turbidity imposed a negative effect on D-dimer by immunonephelometry. The dosage effect testing showed that chylus with 1 200-4 800 turbidity imposed a linear negative effect on D-dimer. Conclusions Triglyceride testing shall be made to chylus samples with >1 200 turbidity and then D-dimer testing results shall be corrected according to linear model, so that the accuracy of results can be obtained.

Key words: Inference test, D-dimer, Immunonephelometry, EPT-A2 document

CLC Number: