Production of candidate secondary reference material for catalytic activity concentration of alpha-amylase in frozen human serum

WANG Jianbing, HAN Liqiao, LIN Haibiao, ZHANG Qiaoxuan, YAN Jun, ZHUANG Junhua, HUANG Xianzhang()   

  1. Department of Clinical Laboratory,Guangdong Provincial Hospital of Chinese Medicine,Guangzhou 510120,Guangdong,China
  • Received:2019-10-14 Online:2021-03-30 Published:2021-03-30
  • Contact: HUANG Xianzhang

摘要:

目的 研制冰冻人血清α-淀粉酶催化活性浓度候选二级标准物质,作为量值传递的标准。方法 参照JJF 1343—2012《标准物质定值的通用原则及统计学原理》和JJF 1644—2017《临床酶学标准物质的研制》要求进行均匀性、稳定性、互换性评价;联合6家通过中国合格评定国家认可委员会(CNAS)认可的校准实验室对标准物质进行定值,评定其均匀性、稳定性和定值引入的不确定度,计算扩展不确定度。结果 标准物质均匀性和稳定性均良好,互通性评价结果显示,4个常规检测系统的测量值均位于95%可信区间内。定值结果:水平1为(106.3±4.3)U/L(k=2),水平2为(268.1±7.9)U/L(k=2)。 结论 研制的冰冻人血清α-淀粉酶催化活性浓度候选二级标准物质标示值可靠,均匀性、稳定性较好,具有互换性。

关键词: α-淀粉酶, 均匀性, 稳定性, 互换性, 标准物质

Abstract:

Objective To develop a candidate secondary reference material for the catalytic activity of alpha-amylase in frozen human serum as a standard for value transmission. Methods According to JJF 1343-2012 General and Statistical Principles for Characterization of Reference Materials and JJF 1644-2017 Production of Reference Materials for Clinical Enzymology,the homogeneity,stability and commutability were evaluated. The value of the reference material was determined by 6 China National Accreditation Service(CNAS) accredited reference laboratories. The uncertainties of the certified values were calculated. Results The results of homogeneity and stability were both good. The measured values were in the 95% confidence interval in the 4 conventional detection systems for commutability. The values were (106.3±4.3)U/L(k=2)on level 1 and (268.1±7.9)U/L(k=2)on level 2. Conclusions The candidate secondary reference material of alpha-amylase has promising homogeneity,stability and commutability,and its labeled values are reliable.

Key words: Alpha-amylase, Homogeneity, Stability, Commutability, Reference material