TOSOHHLC-723G8全自动糖化血红蛋白分析仪性能评价

陈健   

  1. 复旦大学附属华山医院检验科,上海 200040
  • 收稿日期:2013-07-01 出版日期:2013-10-30 发布日期:2013-10-29
  • 作者简介:陈健,男,1973年生,技师,主要从事临床检验工作。

Performance evaluation of TOSOH HLC-723 G8 automatic glycohemoglobin analyzer

CHEN Jian.   

  1. Department of Clinical Laboratory,Huashan Hospital,Fudan University,Shanghai 200040,China
  • Received:2013-07-01 Online:2013-10-30 Published:2013-10-29

摘要:

目的 对以离子交换高效液相色谱法(HPLC)为原理的TOSOH HLC-723 G8全自动糖化血红蛋白分析仪(简称HLC-723 G8)检测全血糖化血红蛋白(HbA1c)的性能进行评价。方法 对HLC-723 G8测定HbA1c的精密度、相关性、线性、携带污染进行评价,并检测54例健康体检者的HbA1c进行参考区间验证。结果 HLC-723 G8测定HbA1c低、高水平批内变异系数(CV)分别为0.00%、0.47%,批间CV分别为0.76%、0.31%;HLC-723 G8与HLC-723 G7全自动糖化血红蛋白分析仪测定HbA1c的结果呈明显相关(r2=0.999,P<;0.001),预期偏倚为0.06,相对偏倚值为0.83%,偏差符合率为100%;在5%~12%范围内线性良好(r2=0.999 6,P<;0.001);高值标本对低值标本的结果无明显携带污染(携带污染率为1.12%);54名健康人群的HbA1c测定结果为4.7%~6.3%,在厂家提供的参考区间内。结论 HLC-723 G8测定HbA1c的性能良好,可在临床使用。

关键词: 高效液相色谱法, 糖化血红蛋白, 性能评价

Abstract:

Objective To evaluate the performance of TOSOH HLC-723 G8 automatic glycohemoglobin analyzer (HLC-723 G8) for the determination of glycosylated hemoglobin A1c (HbA1c) with the principle of ion-exchange high-performance liquid chromatography (HPLC). Methods The precision,correlation,linearity and carryover of HLC-723 G8 were evaluated. The reference interval of HbA1c was verified among 54 healthy subjects. Results The within-run and between-run coefficients of variation (CV) of both low and high levels of HbA1c were 0.00%,0.47% and 0.76%, 0.31%, respectively. HLC-723 G8 and HLC-723 G7 for the determination of HbA1c showed good correlation (r2=0.999,P<0.001). The expected bias was 0.06,the relative bias was 0.83%,and the bias coincidence was 100%. An excellent linearity was found for HbA1c in the range of 5%-12% (r2=0.999 6,P<0.001).There was no obvious carryover in both low and high level samples, and the carryover rate was 1.12%. The levels of HbA1c among the 54 healthy subjects were 4.7%-6.3% in the range provided by manufacturers. Conclusions HLC-723 G8 has a good performance, and can be used in clinical application.

Key words: High-performance liquid chromatography, Glycosylated hemoglobin A1c, Performance evaluation