检验医学 ›› 2013, Vol. 28 ›› Issue (8): 707-710.DOI: 10.3969/j.issn.1673-8640.2013.08.014

• 技术研究与评价.论著 • 上一篇    下一篇

新型碘离子选择性电极法快速测定尿碘评价

王君1,郑静雨2   

  1. 1.武汉科技大学医学院,湖北 武汉 430065;
    2.武汉能斯汉仪器有限责任公司,湖北 武汉 430070
  • 收稿日期:2013-08-30 修回日期:2013-08-30 出版日期:2013-08-30 发布日期:2013-08-30
  • 作者简介:王 君,女,1977年生,硕士,副教授,主要从事功能性食品药品及检测仪器的开发研究。

Assessment of new iodide ion-selective electrode for the determination of urine iodine

WANG Jun1, ZHENG Jingyu2.   

  1. 1.College of Medicine,Wuhan University of Science and Technology,Hubei Wuhan 430065,China;
    2.Wuhan Nengsihan Instrument Co., Ltd., Hubei Wuhan 430070, China
  • Received:2013-08-30 Revised:2013-08-30 Online:2013-08-30 Published:2013-08-30

摘要: 目的 对一种新型碘离子选择性电极快速测定尿碘的方法进行实际应用评价。方法 根据我国机械行业标准JB/T9362-1999《离子选择电极技术条件》要求测定新型碘离子选择性电极的性能参数及选择性系数;以国家标准参考物质(GBW09108~ GBW09110)了解该方法测定的准确度;并应用该方法快速检测126例现场尿样的尿碘含量,并与国家卫生行业标准方法(WS/T107-2006)——砷铈催化分光光度法进行对比试验。结果 新型碘离子选择性电极法的斜率≥55 mV,响应时间<2 min,适用pH值范围为2~10,内阻<500 kΩ,重复性误差<2 mV,稳定性<±8 mV/24 h,检测范围为50~690 μg/L,适用温度范围为5~60 ℃;碘离子对氯离子(Cl-)和硫离子(S2-)等离子的选择性系数达到10-7数量级;尿液中的有机物质均不干扰测定。对国家标准参考物质的测定结果在给定的标准值范围内。2种方法测定126份样本尿碘的结果差异无统计学意义(t=0.527,P=0.604),相关性良好(回归方程为Y=0.972 9X+4.484 5,r=0.978 7)。结论 新型碘离子选择性电极可以用于尿碘的测定。

关键词: 碘, 尿液, 离子选择性电极, 离子选择性系数

Abstract: Objective To assess the application significance of the new iodide ion-selective electrode for the determination of urine iodine. Methods Performance parameters and selectivity coefficients of the new iodide ion-selective electrode were measured according to the national mechanical industry standard JB/T9362-1999 Technical Specifications for the Ion-Selective Electrode. National standard reference materials(GBW09108-GBW09110) were performed for the new iodide ion-selective electrode to evaluate the accuracy. Urine iodine ion concentrations of 126 urine samples were determined by the new method, and the results were compared with the national health care industry standard method-arsenic cerium catalytic spectrophotometric determination(WS/T107-2006). Results The detection range of the new iodide ion-selective electrode was 50-690 μg/L with a slope ≥ 55 mV, the time to response < 2 min, the applicable pH range 2-10, the internal resistance < 500 kΩ, repeatability error < 2 mV, stability <±8 mV/24 h and the range of applicable temperature 5-60 ℃. The selectivity coefficients of the new iodide ion-selective electrode to Cl-, S2- and etc. reached 10-7 grade. The interference of co-existing organic materials was eliminated. The determination results of the national standard reference materials were within the standard value range.There was no statistical significance between the results of urine iodine ion concentrations of 126 urine samples by the new iodide ion-selective electrode and standard methods(t=0.527,P=0.604). There was a good correlation between the 2 methods, and the linear regression equation was Y=0.972 9X+4.484 5(r=0.978 7). Conclusions The new iodide ion-selective electrode can be used for the determination of urine iodine ion.

Key words: Iodine, Urine, Ion-selective electrode, Ion selectivity coefficient

中图分类号: