检验医学 ›› 2026, Vol. 41 ›› Issue (9): 879-883.DOI: 10.3969/j.issn.1673-8640.2026.09.009

• 论著 • 上一篇    下一篇

4种抗磷脂酶A2受体抗体检测方法在诊断原发性膜性肾病中的比较

王均芬, 俞琦, 娄加陶, 范秋灵, 张如霖, 张荣, 孙懿()   

  1. 上海市第一人民医院检验医学中心, 上海 200080
  • 收稿日期:2025-01-17 修回日期:2026-02-26 出版日期:2026-09-30 发布日期:2026-09-29
  • 通讯作者: 孙 懿,E-mail:medsunyi@163.com。
  • 作者简介:王均芬,女,1988年生,硕士,医师,主要从事慢性肾脏病的实验诊断研究。

Comparison of 4 anti-phospholipase A2 receptor antibody determination methods in the diagnosis of primary membranous nephropathy

WANG Junfen, YU Qi, LOU Jiatao, FAN Qiuling, ZHANG Rulin, ZHANG Rong, SUN Yi()   

  1. Department of Clinical Laboratory, Shanghai General Hospital, Shanghai 200080, China
  • Received:2025-01-17 Revised:2026-02-26 Online:2026-09-30 Published:2026-09-29

摘要:

目的 探讨4种方法[酶联免疫吸附试验(ELISA)、间接免疫荧光法(IFA)、流式荧光发光法(FLIA)和化学发光免疫分析法(CLIA)]检测抗磷脂酶A2受体(PLA2R)抗体在原发性膜性肾病(PMN)诊断中的性能。方法 选取2023年9月—2024年6月上海市第一人民医院PMN患者103例(PMN组)、非PMN患者177例(非PMN组)、正常对照者40名(正常对照组)。收集所有研究对象的临床资料,并采用4种方法同时检测血清抗PLA2R抗体。采用受试者工作特征(ROC)曲线得出4种方法检测抗PLA2R抗体诊断PMN的最佳临界值,采用正常对照组血清样本验证临界值。结果 IFA、CLIA、FLIA、ELISA检测血清抗PLA2R抗体的敏感性分别为84.47 %、82.52 %、78.64 %、63.11 %,特异性均为100 %。ELISA、FLIA、CLIA检测血清抗PLA2R抗体诊断PMN的曲线下面积(AUC)分别为0.988 2、0.988 7、0.970 7,临界值分别设定为2.730 RU·mL-1、7.015 U·mL-1、3.705 U·mL-1时,可提升敏感性而不显著降低特异性。IFA的临界值取1∶10时,诊断PMN的效能最高。结论 4种方法中IFA的敏感性最高,CLIA和FLIA次之,ELISA最低。ELISA、FLIA、CLIA的临界值分别取2.730 RU·mL-1、7.015 U·mL-1、3.705 U·mL-1时,可提高PMN的诊断敏感性而不显著降低特异性。

关键词: 抗磷脂酶A2受体抗体, 间接免疫荧光法, 流式荧光发光法, 化学发光免疫分析法, 原发性膜性肾病

Abstract:

Objective To investigate the performance of 4 methods [enzyme-linked immunosorbent assay (ELISA),indirect immunofluorescence assay(IFA),flow luminescence immunoassay(FLIA)and chemiluminescence immunoassay(CLIA)] in determining anti-phospholipase A2 receptor(PLA2R)antibodies in the diagnosis of primary membranous nephropathy(PMN). Methods A total of 103 patients with PMN(PMN group),177 patients without PMN(non-PMN group)and 40 healthy subjects(healthy control group)were enrolled from Shanghai General Hospital from September 2023 to June 2024. The clinical data were collected,and serum anti-PLA2R antibodies were determined using the 4 methods. The optimal cut-off value for determining anti-PLA2R antibodies in diagnosing PMN was obtained using receiver operating characteristic(ROC)curve,and the cut-off value was verified using the serum samples of healthy control group. Results The sensitivities of IFA,CLIA,FLIA and ELISA for determining serum anti-PLA2R antibodies were 84.47%,82.52%,78.64% and 63.11%,respectively. The specificities were 100%. The areas under curves(AUC)for determining serum anti-PLA2R antibodies in diagnosing PMN by ELISA,FLIA and CLIA were 0.988 2,0.988 7 and 0.970 7,respectively. The cut-off values were set at 2.730 RU·mL-1, 7.015 U·mL-1 and 3.705 U·mL-1,respectively,which could improve the sensitivity without significantly reducing the specificity. The cut-off value of IFA was 1∶10,and its diagnostic efficacy for PMN was the highest. Conclusions Among the 4 methods,IFA has the highest sensitivity,followed by CLIA and FLIA,and ELISA has the lowest. When the cut-off values of ELISA,FLIA and CLIA are set at 2.730 RU·mL-1,7.015 U·mL-1 and 3.705 U·mL-1,respectively,the diagnostic sensitivity of PMN could be improved without significantly reducing specificity.

Key words: Anti-phospholipase A2 receptor antibody, Indirect immunofluorescence assay, Flow luminescence immunoassay, Chemiluminescence immunoassay, Primary membranous nephropathy

中图分类号: