BD FACSCantoⅡ流式细胞仪方法学性能验证

陈淑英1, 陈健1, 林勇1, 时朝辉2   

  1. 1.复旦大学附属华山医院检验医学科,上海200040
    2.贵州医科大学,贵州 贵阳550004
  • 收稿日期:2015-04-29 出版日期:2016-05-30 发布日期:2016-06-13
  • 作者简介:null

    作者简介:陈淑英,女,1989年生,硕士,技师,主要从事神经免疫学研究。

    通讯作者:林 勇,联系电话:021-52888321。

Methodology verification on the performance of BD FACSCantoⅡflow cytometry analyzer

CHEN Shuying1, CHEN Jian1, LIN Yong1, SHI Chaohui2   

  1. 1.Department of Clinical Laboratory,Huashan Hospital,Fudan University,Shanghai 200040,China
    2.Guizhou Medical University,Guiyang 550004,Guizhou,China
  • Received:2015-04-29 Online:2016-05-30 Published:2016-06-13

摘要:

目的 对BD FACSCantoⅡ流式细胞仪的主要分析性能进行方法学验证。方法 参照美国临床实验室标准化协会(CLSI)系列文件和相关参考资料所提供的方案,在BD FACSCantoⅡ流式细胞仪上对科室常用的流式检测项目(外周血淋巴细胞亚群:CD3+T细胞、CD3+CD4+T细胞、CD3+CD8+T细胞、B细胞和NK细胞)的精密度、准确度、线性范围和参考区间进行验证,将其结果与生产商声明的性能指标进行比较。结果 BD FACSCantoⅡ流式细胞仪检测外周血淋巴细胞亚群的精密度、准确度、线性范围和参考区间均与生产商所示的性能参数相符。结论 BD FACSCantoⅡ 流式细胞仪检测外周血淋巴细胞亚群的方法学性能良好,测量结果准确、可靠且重复性好,能够满足临床的需要。

关键词: 流式细胞仪, 外周血淋巴细胞亚群, 性能验证

Abstract:

Objective To verify the main performance of BD FACSCantoⅡflow cytometry analyzer. Methods According to the Clinical and Laboratory Standards Institute(CLSI)documents and projects provided from other literatures,the analytical performance of BD FACSCantoⅡflow cytometry analyzer was evaluated by precision,accuracy,linearity and reference range of common test items(peripheral blood lymphocyte subtypes:CD3+T cell,CD3+CD4+T cell,CD3+CD8+T cell,B cell and NK cell). The results were compared with the claims of manufacturers. Results The precision,accuracy,linearity and reference range of BD FACSCantoⅡflow cytometry analyzer in the determination of peripheral blood lymphocyte subtypes were in line with the performance parameters of manufacturers. Conclusions BD FACSCantoⅡflow cytometry analyzer has a good performance,accurate measurement and a high repeatability in the determination of peripheral blood lymphocyte subtypes,which can satisfy the need in clinics.

Key words: Flow cytometry analyzer, Peripheral blood lymphocyte subtypes, Performance verification