上海地区细菌鉴定项目室间质量评价的23年回顾

陈蓉, 徐蓉, 刘学杰, 葛平   

  1. 上海市临床检验中心,上海 200126
  • 收稿日期:2015-06-26 出版日期:2015-12-20 发布日期:2016-01-04
  • 作者简介:null

    作者简介:陈 蓉,女,1979年生,学士,主管技师,主要从事微生物检测和质量管理工作。

  • 基金资助:
    上海市卫生和计划生育委员会资助项目(2013ZYJB0010)

Review on external quality assessment of bacterial ide.pngication items in Shanghai for past 23 years

CHEN Rong, XU Rong, LIU Xuejie, GE Ping.   

  1. Shanghai Center for Clinical Laboratory, Shanghai 200126, China
  • Received:2015-06-26 Online:2015-12-20 Published:2016-01-04

摘要:

目的 了解上海市医疗机构近23年来微生物室间质量评价(EQA)细菌鉴定项目的检测质量。方法 首先对EQA活动的频率、分组、质控品性状、运输模式、回报方式的演变进行分析;收集1992至2014年参加上海市临床检验中心(SCCL)细菌鉴定EQA项目的反馈结果,并对主要的14个菌属的细菌鉴定符合率进行统计分析。结果 EQA活动的频率由4次/年减少到2次/年,样本数量由3个/次增加到5个/次;由最初的不分组到现在的分组统计;2007年之前质控品为半固体穿刺状,2007年后采用冷冻干燥技术制成冻干品;运输模式由医务人员上门自取转变成冷链系统的全程监控;回报方式由纸质回报到目前的网络上传。上海市医疗机构实验室参加SCCL微生物EQA活动的数量由最初的80家逐年增加到目前的134家,增加了约68%;每年的合格率(≥80分为合格)和优秀率(满分100分为优秀)也在逐年增加,至2014年合格率已达到100%,优秀率达到90%。发放的菌株涵盖23个属105个种,其中的葡萄球菌属、肠球菌属、肠杆菌科鉴定符合率较高,符合率多次达到100%。结论 EQA活动形式的变化使得质量评价工作更加公正、规范。上海市医疗机构的细菌鉴定质量在逐年提高,可更好地为临床提供精准的检验结果。

关键词: 微生物, 室间质量评价, 质量管理

Abstract:

Objective To study the determination quality of microbiology external quality assessment (EQA) of bacterial ide.pngication items in Shanghai medical institutions for past 23 years.Methods Firstly, the EQA activity frequency, grouping, quality control material characteristic, transportation mode and returning way were analyzed. The results of microbiology bacterial ide.pngication items among participating laboratories in EQA of Shanghai Center for Clinical Laboratory (SCCL) from 1992 to 2014 were collected and analyzed statistically. The coincidence for the determinations of 14 major bacteria was analyzed statistically. Results EQA activity frequency reduced from 4 times per year to 2 times per year, and the number of samples increased from 3 samples per time to 5 samples per time. Subjects were not classified before, and now subjects did grouping. Before 2007, quality control materials were puncture-like semi-solid, after that, quality control materials were treated by freeze-drying technology as lyophilized form. Transportation modes were from taking by medical personnels themselves to cold chain system delivery in order to monitor whole process. Returning way was from paper to web uploading. The number of participating laboratories increased from 80 to 134 now, and the increasing rate was about 68%. The pass rate (score ≥80) and the excellent rate (score=100) increased year by year. The pass rate in 2014 was 100%, and the excellent rate in 2014 was 90%. The participating isolates were 23 genera with 105 kinds, and the coincidence rates of Staphylococcus, Enterococcus and Enterobacteriaceae ide.pngication were high, and many of them reached 100%. Conclusions The change of EQA activity mode makes it more equitable and standard for quality assessment. The microbiology bacterial ide.pngication quality increases year by year in Shanghai medical institutions, providing accurate results for clinic.

Key words: Microorganism, External quality assessment, Quality management