Laboratory Medicine ›› 2021, Vol. 36 ›› Issue (5): 549-553.DOI: 10.3969/j.issn.1673-8640.2021.05.018

Previous Articles     Next Articles

Establishment and preliminary evaluation of HbA1c determination harmonization plan of Xuhui District of Shanghai

WANG Wei1, WANG Lei1(), WANG Beili2, ZHOU Fangfang1, HU Jun3, WU Pei4, YIN Fen5, YUAN Kaiying6, HOU Yumin7, DIAO Duanqin8, LI Guoying9, CHEN Min10, GAO Yuhong11   

  1. 1. Department of Clinical Laboratory,Shanghai Eighth People's Hospital,Shanghai 200235,China
    2. Department of Clinical Laboratory,Zhongshan Hospital,Fudan University,Shanghai 200032,China
    3. Community Healthcare Center of Xietu Street,Shanghai 200025,China
    4. Community Healthcare Center of Lingyun Street,Shanghai 200237,China
    5. Community Healthcare Center of Hongmei Street,Shanghai 200233,China
    6. Community Healthcare Center of Kangjian Street,Shanghai 200235,China
    7. Community Healthcare Center of Fenglin Street,Shanghai 200032,China
    8. Community Healthcare Center of Changqiao Street,Shanghai 200231,China
    9. Community Healthcare Center of Caohejing,Shanghai 200233,China
    10. Community Healthcare Center of Longhua,Shanghai 200232,China
    11. Community Healthcare Center of Huajing Town,Shanghai 200231,China
  • Received:2020-06-12 Online:2021-05-30 Published:2021-05-30
  • Contact: WANG Lei

Abstract:

Objective To establish the harmonization plan of glycated hemoglobin A1c(HbA1c) determination in Xuhui District of Shanghai,and to improve the consistency of HbA1c determination results in Xuhui community healthcare centers. Methods Shanghai Eighth People's Hospital,the leading unit of the HbA1c determination harmonization plan in Xuhui District,regularly prepared fresh mixed whole blood samples. The target values were determined by the laboratory of Zhongshan Hospital affiliated to Fudan University,which was certified as the second level reference laboratory of the National Glycohemoglobin Standardization Program(NGSP). The samples were distributed to the 9 laboratories of community healthcare centers participating in the HbA1c determination harmonization plan in Xuhui District. HbA1c was determined by high-performance liquid chromatography(HPLC) in each laboratory. The results of 10 laboratories,including the leading unit,were compared,and the pass rate,deviation and coefficient of variation(CV) of HbA1c determination results were analyzed. Results A total of 5 batches of comparison samples were distributed. The pass rates of 5 comparisons in 10 laboratories were 100%,10%,100%,100% and 50%,respectively. The bias ranges were -4.44%-5.53%,-10.39%-3.39%,-5.00%-7.41%,-4.44%-4.65% and -5.13%-10.91%,and the CV were 1.26%-2.71%,1.20%-4.07%,0.89%-3.36%,0.63%-1.90% and 2.14%-3.62%,respectively. Conclusions By participating in the HbA1c determination harmonization plan of Xuhui District,the consistency of HbA1c determination results in the laboratories of community healthcare centers has been improved.

Key words: Glycated hemoglobin, Harmonization, Diabetes mellitus