Application of candidate reference method of apolipoprotein A1 and apolipoprotein B in external quality assessment

LI Qing, JU Yi(), TANG Liping, JIN Zhonggan, SUN Hewei   

  1. Shanghai Center for Clinical Laboratory,Shanghai 200126,China
  • Received:2020-07-19 Online:2021-03-30 Published:2021-03-30
  • Contact: JU Yi

Abstract:

Objective To investigate the application of candidate reference method of apolipoprotein(apo) A1 and apo B in external quality assessment(EQA). Methods Five levels of frozen human serum products were used for EQA samples. Immunoturbidimetry (ITA)was used for the detection. EQA samples were quantified by isotope dilution liquid chromatography-tandem mass spectrometry(ID-LC-MS/MS). Samples underwent denaturing,alkalization and tryps in digestion,and were then separated by C18 column and quantified by positive ion mode and multiple reaction monitoring. Results ITA was used in 340 participants. As for apo A1,ID-LC-MS/MS gave values lower than the median of ITA for EQA2-EQA4,but not for EQA1 and EQA5. As for apo B,except EQA4,all ID-LC-MS/MS values were lower than the median of ITA. Conclusions Candidate reference method based on ID-LC-MS/MS can offer application value of assessment based on target value.

Key words: Apolipoprotein A1, Apolipoprotein B, Candidate reference method, External quality assessment