Laboratory Medicine ›› 2021, Vol. 36 ›› Issue (10): 1064-1069.DOI: 10.3969/j.issn.1673-8640.2021.010.016

Previous Articles     Next Articles

Isotope dilution liquid chromatography tandem mass spectrometry candidate reference method for serum testosterone

SUN Hewei, JU Yi(), ZHU Lingfeng, LI Qing, JIN Zhonggan, ZHANG Sujie   

  1. Shanghai Center for Clinical Laboratory,Shanghai 200126,China
  • Received:2020-09-15 Online:2021-10-30 Published:2021-11-08
  • Contact: JU Yi

Abstract:

Objective To establish a candidate reference method for the determination of human serum testosterone based on liquid chromatography tandem mass spectrometry(LC-MS/MS),and to use it into assessing the trueness of clinical method(microparticle chemiluminescence immunoassay). Methods Testosterone-2,3,4-13C3 was added as internal standard. Serum samples were treated with hexane-ethyl acetate(50∶50,volume fraction). The upper layer was dried with nitrogen flow. The residual were reconstituted with mobile phase and analyzed by LC-MS/MS. Serum testosterone concentration was calculated by 5-point bracketing method. The isotope dilution liquid chromatography tandem mass spectrometry(ID-LC-MS/MS)(carryover,specificity,linearity,imprecision,accuracy) was validated according to the Clinical and Laboratory Standards Institute(CLSI) C62-A and EP15-A3. The trueness of microparticle chemiluminescence immunoassay was evaluated according to CLSI EP9-A3. Results The ID-LC-MS/MS with 5 min total run time and without carryover was validated over a concentration range of 0.034-22.000 ng/mL,and the specificity was good. The within-run imprecision was ≤2.3%,and the between-run imprecision was ≤2.2%. The bias of analyzing the candidate reference material SRM971(Level male)was 0.2%. The relative biases of analyzing 2017 RELA-A and 2017 RELA-B were -2.9% and -3.3%,respectively. The uncertainty for 2017 RELA-A was 0.11 ng/mL. The linearity between ID-LC-MS/MS and microparticle chemiluminescence immunoassay was good(r=0.963). However,its average relative bias was large,which total average relative bias was -24.4%,the average relative bias of low concentration group(≤1 ng/mL) was -62.4%,and the average relative bias of high concentration group(>1 ng/mL) was 6.3%. Conclusions The ID-LC-MS/MS established successfully for serum testosterone has good precision and accuracy,with simple preparation and short analysis time and may be used as a candidate reference method.

Key words: Testosterone, Isotope dilution liquid chromatography tandem mass spectrometry, Methodological assessment, Candidate reference method

CLC Number: