Laboratory Medicine ›› 2023, Vol. 38 ›› Issue (8): 776-780.DOI: 10.3969/j.issn.1673-8640.2023.08.013

Previous Articles     Next Articles

Evaluation on the consistency of blind sample test results of 9 SARS-CoV-2 nucleic acid determination kits in Shanghai,China

YANG Xue, ZHU Jun, JIANG Lingli, WANG Qing(), HU Xiaobo()   

  1. Shanghai Center for Clinical Laboratory,Shanghai 200126,China
  • Received:2022-08-25 Revised:2022-11-17 Online:2023-08-30 Published:2023-10-30

Abstract:

Objective To know the consistency of severe acute respiratory syndrome coronavirus 2(SARS-CoV-2) nucleic acid determination results of clinical laboratories in Shanghai,China. Methods The results of 4 SARS-CoV-2 nucleic acid blind sample tests and the cycle threshold(Ct)value of positive results in clinical laboratories were analyzed. Results Totally,249 clinical laboratories participating in the 4 blind sample tests used 9 kinds of determination kits to determine 2 negative blind samples and 3 positive blind samples. The correct rate of negative samples and the determination rate of positive samples were all 100%. The difference ranges of Ct values' x of ORF1ab gene and N gene in 3 positive samples of 9 determination kits were 2.58-2.66 and 2.88-3.06. The difference ranges of range(R) were 7.98-8.15 and 7.71-7.73,respectively. The coefficient of variation(CV) of Ct values of ORF1ab gene and N gene in 5 kit groups were all <5%. Only one kit results showed large variation within and between groups. Conclusions The determination capabilities of SARS-CoV-2 nucleic acid determination among clinical laboratories in Shanghai could all meet the requirements,but the Ct values of different determination kits are different. When interpreting the results based on the Ct value,false negative or positive results might occur,which should be paid attention by clinical laboratories and relevant management institutes.

Key words: Severe acute respiratory syndrome coronavirus 2, Nucleic acid determination, Blind sample, Cycle threshold, External quality assessment

CLC Number: