Laboratory Medicine ›› 2016, Vol. 31 ›› Issue (4): 293-298.DOI: 10.3969/j.issn.1673-8640.2016.04.011

• Orginal Article • Previous Articles     Next Articles

Comparison on the performances of 5 rapid systems for the determination of C-reactive protein

JIANG Lingli1, TANG Dahai2, ZHANG Jian1, ZHANG Fan3, GE Danhong1, XIAO Yanqun1, ZHU Lingfeng1   

  1. 1. Shanghai Center for Clinical Laboratory, Shanghai 200126, China
    2. No. 411Hospital of Chinese People's Liberation Army, Shanghai 200081, China
    3. Nanjing Brain Hospital Affiliated to Nanjing Medical University, Nanjing 210029, Jiangsu, China
  • Received:2015-10-08 Online:2016-04-30 Published:2016-05-12

Abstract:

Objective To evaluate the performances of 5 common systems for C-reactive protein(CRP) determination. Methods The imprecision [coefficient of variation(CV)],accuracy,linearity and anti-interference ability of 5 systems for CRP determination were evaluated. The 5 systems included Uppergold U2 gold spot method system,NEPHSTAR specific protein analyzer,ABX Micros CRP200 automatic blood analyzer,i-CHROMA Reader immune fluorescence analyzer and QuikRead 101 detecting system. Their original reagents were used. The 5 systems were labelled as A,B,C,D and E in turn. The reference range was validated. The results were compared with those of Dade Behring BN ProSpec automatic system (DADE). Results The within-run and between-run precisions of the 5 systems were 1.01%-16.91%,which met the requirements from manufacturers,but they were lower than those of DADE. Passing-Bablok regression analysis showed that the regression equation slopes between DADE and the 5 systems were 0.88-1.73. The correlations were good between DADE and A,B,C systems,but the correlations were poor between DADE and D,E systems. For 50 mg/L CRP as cut-off value,the consistency of the 5 systems for CRP determination with DADE was 92.4%-100%. The linear range of 4 systems was 5-160 mg/L,and the linear range of 1 system was 5-100 mg/L. The interference rates of 1 500 FTU chyle,0.2 mg/L free bilirubin(FBil),0.2 mg/L conjugated bilirubin(CBil)and 5 g/L hemoglobin(Hb) were <10%. The reference value met the requirements from manufacturers. Conclusions The performances of the 5 systems are good. However,there is a bias between the 5 systems and DADE system,which should be considered in clinic.

Key words: C-reactive protein, Rapid determination method, Immunoturbidimetric assay, Immunonephelometric assay