Evaluation of uncertainty of serum total bilirubin in calibration curve method by Monte Carlo method

MENG Shuting1, ZHOU Weiping2, DAI Huifen3, DONG Deping1, DONG Yefeng1, GE Renmei1, WANG Huimin4()   

  1. 1. Department of Clinical Laboratory,Hai'an Hospital of Nantong University,Nantong 226600,Jiangsu,China
    2. Department of Clinical Laboratory,Tongzhou People's Hospital,Nantong 226300,Jiangsu,China
    3. Department of Clinical Laboratory,Traditional Chinese Medical Hospital of Hai'an County,Nantong 226600,Jiangsu,China
    4. Department of Clinical Laboratory,the Affliated Hospital of Nantong University,Nantong 226601,Jiangsu,China
  • Received:2021-12-19 Revised:2022-03-05 Online:2022-07-30 Published:2022-08-26
  • Contact: WANG Huimin

Abstract:

Objective To evaluate the uncertainty process for total bilirubin reference measurement procedure which was recommended by the Joint Committee for Traceability in Laboratory Medicine(JCTLM) used by the Monte Carlo method(MCM) through optimizing. To realize quantity traceability about the uncertainty evaluation of serum total bilirubin under reference laboratory conditions. Methods Applying Doumas method which was recommended by JCTLM,sample A and sample B from external quality assessment scheme for reference laboratories of the International Federation of Clinical Chemistry and Laboratory Medicine(IFCC) in 2014 were determined. Based on software MATLAB,the uncertainty of total bilirubin determination was evaluated by MCM. Results The uncertainty model which was evaluated by the Guide to the Expression of Uncertainty in Measurement(GUM)was established. Parameters related to the calibration curve were as follows:slope(b) was 0.007 531 6,the standard deviation of slope was 0.000 025 51,mean concentration was 87.284 4 mg/L,the standard deviation of mean concentration was 0.196 5,mean absorbance($\bar{y}$) was 0.657 917,the standard deviation of mean absorbance was 0.000 365 4,intercept(a) was 0.000 525 16,and the standard deviation of intercept was 0.002 699. The results of total bilirubin determination under laboratory conditions were as follows:the concentration of sample A x was 39.368 8 mg/L,the standard uncertainty of sample A synthesis(u) was 0.603 2,95% confidence interval(CI) was 38.188 0-40.553 5,the concentration of sample B x was 18.703 1 mg/L,the standard uncertainty of sample B synthesis(u)was 0.296 4,and 95%CI was 18.122 5-19.284 4. Output quantity of MCM presented skewed normal distribution,showing peak fat-tail. The GUM uncertainty framework(GUF) inclusion interval was verified with adaptive MCM. The absolute deviation dlow and dhigh values of the 2 groups were greater than the numerical tolerance 0.05. The results of GUF did not pass the MCM verification. Conclusions With the help of visual software MATLAB,MCM can optimize calculation process,improve the accuracy of inspection results highly compared with the classical GUF.

Key words: Monte Carlo method, Uncertainty, Total bilirubin, Standard curve