Evaluation on measurement uncertainty of serum total bilirubin concentration by reference procedure

DAI Huifen1, WANG Huimin2, WANG Jianxin2, JI Huoyan2, MENG Shuting2   

  1. 1. Department of Clinical Laboratory,Traditional Chinese Medical Hospital of Haian County,Haian 226600,Jiangsu,China
    2. Department of Clinical Laboratory,Affiliated Hospital of Nantong University,Nantong 226001,Jiangsu,China
  • Received:2015-11-15 Online:2016-05-30 Published:2016-06-13

Abstract:

Objective To investigate the evaluation method for measurement uncertainty of serum total bilirubin (TBil) concentration in clinical reference laboratories.Methods The International Federation of Clinical Chemistry and Laboratory Medicine(IFCC) external quality assessment scheme for reference laboratories in laboratory medicine(RELA)sample was measured by Doumas serum TBil reference procedure of the Joint Committee for Traceability in Laboratory Medicine(JCTLM),and the sources of uncertainty components were identified. The evaluation of uncertainty components was processed,and then a combined standard uncertainty was calculated. Finally,an expanded uncertainty was obtained.Results The TBil average concentration of 2013 RELA-A sample was 52.30 μmol/L,the combined standard uncertainty was 0.67 μmol/L,the 95% confidence interval was taken,and the coverage factor(k)was 2. The expanded uncertainty was 1.34 μmol/L. The TBil average concentration of 2013 RELA-B sample was 81.60 μmol/L,the combined standard uncertainty was 0.70 μmol/L,the 95% confidence interval was taken,and the k was 2. The expanded uncertainty was 1.40 μmol/L.Conclusions The established method can meet the requirements of clinical laboratories for serum TBil reference procedure.

Key words: Total bilirubin, Serum, Reference procedure, Measurement uncertainty evaluation

CLC Number: