Laboratory Medicine ›› 2022, Vol. 37 ›› Issue (2): 108-111.DOI: 10.3969/j.issn.1673-8640.2022.02.002

Previous Articles     Next Articles

Evaluation of the interference of serum index in the determination of prealbumin by immunoscattering turbidimetry according to CLSI EP07-A3 document

HAN Guang1, XU Qiongfeng2, DUAN Shumin3, LUO Jiahuan4, WANG Yunxiu1, ZHANG Xiujuan1, WAN Zemin1, KE Peifeng1, HUANG Xianzhang1()   

  1. 1. Guangdong Provincial Hospital of Chinese Medicine,Guangzhou 510120,Guangdong,China
    2. Aviation Health Center of China Southern Airlines Co.,Ltd.,Guangzhou 510400,Guangdong,China
    3. Jinshazhou Hospital of Guangzhou University of Traditional Chinese Medicine,Guangzhou 510168,Guangdong,China
    4. Sun Yat-sen Memorial Hospital,Sun Yat-sen University,Guangzhou 510120,Guangdong,China
  • Received:2020-06-23 Revised:2021-07-27 Online:2022-02-28 Published:2022-03-15
  • Contact: HUANG Xianzhang

Abstract:

Objective To investigate the interference effect of hemolysis,icterus and lipemia on the determination of prealbumin(PA) by immunoscattering turbidimetry. Methods A fresh serum sample was used as a basic sample,interfering substances [hemoglobin(Hb),bilirubin and lipid] were added,and the PA level was determined by immunoscattering turbidimetry. According to the EP07-A3 document of the Clinical and Laboratory Standards Institute(CLSI),the paired difference experiment and dose-effect experiment were performed. Results PA had no interference with 10 g/L Hb(hemolysis index of 1 000)and 200 mg/L(340 μmol/L) bilirubin(icterus index of 20). The lipid turbidity interfering substance with a lipid turbidity index of 987 had a negative interference on the determination of PA. The degree of interference had a linear relationship with the lipid turbidity index. For samples with low values of PA,when the lipid turbidity index≤120,there was no interference to PA determination. The linear equation between lipid turbidity index and interference rate was Y=-0.077X+0.979(r 2=0.999). For samples with high values of PA,when the lipid turbidity index≤203,there was no interference to PA determination. The linear equation between lipid turbidity index and interference rate was Y=-0.040X-0.168(r2=0.992). Conclusions Generally clinical hemolytic and icteric samples will not interfere with the PA determination results by immunoscattering turbidimetry. The samples with lipid turbidity index>120 will cause negative interference to PA determination results.

Key words: Prealbumin, Immunoscattering turbidimetry, Serum index, Interference

CLC Number: