Simultaneous determination of 5 steroid hormones in human serum by ultra-performance liquid chromatography-tandem mass spectrometry and establishment of reference intervals in healthy adults

LU Youli, YANG Shuangshuang, ZHANG Meiwei, OU Meixian, DONG Chunxia, SHEN Weiwei, JIANG Fengli, LI Shuijun()   

  1. Shanghai Engineering Research Center of Phase Ⅰ Clinical Research and Quality Consistency Evaluation for Drugs,Shanghai Institute of Clinical Mass Spectrometry,Shanghai Xuhui Central Hospital,Zhongshan-Xuhui Hospital Affiliated to Fudan University,Shanghai 200031,China
  • Received:2021-07-02 Revised:2021-08-29 Online:2022-02-28 Published:2022-03-15
  • Contact: LI Shuijun

Abstract:

Objective To establish and validate a method for the simultaneous determination of testosterone(T),dihydrotestosterone(DHT),androstenedione(AD),dehydroepiandrosterone(DHEA) and 17alpha-hydroxyprogesterone(17OHP) in human serum by ultra-performance liquid chromatography-tandem mass spectrometry(UPLC-MS/MS),and to establish the reference intervals of 5 steroid hormones in healthy adults. Methods The serum samples were purified by SOLAμ HRP solid-phase extraction,and they were separated by Poroshell 120 EC-C18 analytical column. Mobile phase was constituted of 30% acetonitrile solution and 90% acetonitrile solution containing 0.05% formic acid with gradient elution. The electrospray positive ion multi-reaction monitoring mode was used for determination,and the isotope internal standard method was used to quantify the concentration of steroid hormones. The performance of the method was verified comprehensively,and the reference intervals suitable for the local population in Shanghai were established. Results The minimum limits of quantification of the 5 steroid hormones in human serum were 0.01-0.10 ng/mL. The standard curve range covered 1 000 times (r2≥0.999 5) and showed good linearity,and the sample after dilution can be reported to cover the range of 5 000 times. The average accuracy of 5 steroid hormones ranged from 95.0% to 105.6%,with the precision of <8%. None of the 8 structural analogs interfered with the target analytes,high triglyceride blood did not affect the determination results,and hemolysis resulted in low DHEA concentration. The extraction recoveries of the analytes were 60.5%-93.3%,and there was no obvious matrix effect after calibration of internal standard. Analytes demonstrated good stability at room temperature for 4 h,at -25--15 ℃ for 30 d or at -90--70 ℃ for 100 d. The method was used and participated in a survey of endocrine accuracy verification supplied by the National Center for Clinical Laboratories in 2019. The determination deviations of T and 17OHP were -5.00%-0.38%. The results of 367 physical examination subjects were analyzed and showed that there was statistical significance in the concentration of 5 steroid hormones between females and males(P<0.05). There was statistical significance between the 2 age subgroups(aged 16-50 years and aged>50 years) of female groups(P<0.05). Conclusions A UPLC-MS/MS has been established for the determination of 5 steroid hormones in human serum,and the reference intervals suitable for the Central Laboratory of Shanghai Xuhui Central Hospital have been verified.

Key words: Ultra-performance liquid chromatography, Mass spectrometry, Steroid hormone, Human serum, Reference interval