Laboratory Medicine ›› 2020, Vol. 35 ›› Issue (8): 811-817.DOI: 10.3969/j.issn.1673-8640.2020.08.016

Previous Articles     Next Articles

Establishment and performance evaluation of candidate reference measurement procedure for serum aldosterone by liquid chromatography-tandem mass spectrometry

ZOU Jihua1, LI Quanle1, SHEN Min1, ZHANG Man2, YANG Xiaodong1, TU Minmin1, FENG Dongfang3, SUN Hongna1   

  1. 1. Reference Laboratory,Medical System Biotechnology Co.,Ltd.,Ningbo 315104,Zhejiang,China
    2. Department of Clinical Laboratory,Beijing Shijitan Hospital,Capital Medical University,Beijing 100038,China
    3. Department of Clinical Laboratory,General Hospital of Tibet Military Region,Lhasa 850007,Tibet,China
  • Received:2020-02-13 Online:2020-08-30 Published:2020-09-24

Abstract:

Objective To establish a candidate reference measurement procedure for serum aldosterone by isotope dilution liquid chromatography-tandem mass spectrometry(ID-LC-MS/MS). Methods The sample was prepared by liquid-liquid extraction with methyltert-bytylether as extractant. The aldosterone-2H7 was selected as internal standard. Samples were prepared by liquid-liquid extraction,and aldosterone was detected by ID-LC-MS/MS in negative ion mode. The basic analysis performance of the established candidate reference measurement procedure,including accuracy,precision,sensitivity and linearity,was evaluated. Results The linear range of aldosterone was 0.08-3.55 nmol/L. The limit of detection and the limit of quantitation were (LOQ)0.012 nmol/L and 0.045 nmol/L,respectively. The within-run coefficients of variation(CV) was <5.0%,and the between-run CV was <4.1%. The average recovery rates were 98.52%-100.14%. The bias of the International Federation of Clinical Chemistry and Laboratory Medicine External Quality Assessment Scheme for Reference Laboratories in Laboratory Medicine(RELA) was <1.6%. Conclusions The established ID-LC-MS/MS for detecting aldosterone in serum is accuracy and precise,and may be used as a candidate reference measurement procedure.

Key words: Aldosterone, Isotope dilution liquid chromatography-tandem mass spectrometry, Reference method

CLC Number: