Accuracy evaluation of routine measurement procedure for serum homocysteine based on liquid chromatography-tandem mass spectrometry

ZHONG Qiaoling1, SUN Ji2, WENG Xuqi3, FENG Ping3, XIE Danyun3   

  1. 1. Department of Clinical Laboratory,the Second People's Hospital of Banan District,Chongqing 400054,China
    2. Technique Center,Ningbo Entry-Exit Inspection and Quarantine Bureau,Ningbo 315000,Zhejiang,China
    3. Reference Laboratory,Medical System Biotechnology Co.,Ltd.,Ningbo 315104,Zhejiang,China
  • Received:2018-03-20 Online:2018-08-10 Published:2018-09-07

Abstract:

Objective To evaluate the accuracy of enzymatic cycling method for serum homocysteine(Hcy) based on isotope-dilution liquid chromatography-tandem mass spectrometry(ID-LC/MS/MS) as a contrast method.Methods A total of 20 fresh serum samples were determined by enzymatic cycling method and ID-LC/MS/MS,and the results were compared. The levels of serum Hcy were adjusted until the results were consistent,according to the results of 5 working calibrations,which were determined by enzymatic cycling method and ID-LC/MS/MS.Results The results of standard reference material(SRM) 1950 determined by ID-LC/MS/MS were consistent with certificate value,and the relative bias was <1.0%. The linear equation of ID-LC/MS/MS and enzymatic cycling method was Y=1.204 1X+0.364 6(r2=0.987 9),and the average bias was 24.0%. After adjusting,the linear equation was Y=0.988 6X+0.260 3(r2=0.999 7),and the average bias was -0.09%.Conclusions The levels of serum Hcy determined by ID-LC/MS/MS are accurate,which can be used for the accuracy evaluation of routine measurement procedure for serum Hcy.

Key words: Homocysteine, Isotope-dilution liquid chromatography-tandem mass spectrometry, Enzymatic cycling method, Accuracy evaluation