Laboratory Medicine ›› 2015, Vol. 30 ›› Issue (5): 416-420.DOI: 10.3969/j.issn.1673-8640.2015.05.003
Previous Articles Next Articles
SONG Binbin, QIN Jiaqian, PENG Yingfei, ZHANG Chunyan, WU Jiong, WANG Beili, GUO Wei, PAN Baishen
Received:
2014-04-03
Online:
2015-05-30
Published:
2015-06-17
CLC Number:
SONG Binbin, QIN Jiaqian, PENG Yingfei, ZHANG Chunyan, WU Jiong, WANG Beili, GUO Wei, PAN Baishen. Establishment on a LC-MS/MS and its performance characteristic evaluation for the determination of serum 25-hydroxyvitamin D[J]. Laboratory Medicine, 2015, 30(5): 416-420.
Add to citation manager EndNote|Ris|BibTeX
URL: https://www.shjyyx.com/EN/10.3969/j.issn.1673-8640.2015.05.003
物质 | 母离子>子离子 (m/z) | 锥孔电压 (kV) | 碰撞能量 (eV) |
---|---|---|---|
25(OH)D2 | 413.35>355.30 | 16.00 | 8.00 |
25(OH)D2-d3 | 416.35>358.30 | 16.00 | 8.00 |
25(OH)D3 | 401.35>159.10 | 16.00 | 28.00 |
25(OH)D3-d6 | 407.35>159.10 | 16.00 | 28.00 |
物质 | 母离子>子离子 (m/z) | 锥孔电压 (kV) | 碰撞能量 (eV) |
---|---|---|---|
25(OH)D2 | 413.35>355.30 | 16.00 | 8.00 |
25(OH)D2-d3 | 416.35>358.30 | 16.00 | 8.00 |
25(OH)D3 | 401.35>159.10 | 16.00 | 28.00 |
25(OH)D3-d6 | 407.35>159.10 | 16.00 | 28.00 |
分析物 | 水平 | 理论值 | 测定值1 | 测定值2 | s | CV (%) | 偏倚(%) | |
---|---|---|---|---|---|---|---|---|
25(OH)D2 | 1 | 500.00 | 484.50 | 515.50 | 500.00 | 21.92 | 4.40 | 0.00 |
2 | 250.00 | 247.50 | 249.75 | 248.63 | 1.59 | 0.60 | -0.60 | |
3 | 125.00 | 129.50 | 124.00 | 126.75 | 3.89 | 3.10 | 1.40 | |
4 | 62.50 | 61.75 | 68.00 | 64.88 | 4.42 | 6.80 | 3.80 | |
5 | 25.00 | 22.00 | 21.75 | 21.88 | 0.18 | 0.80 | -12.50 | |
6 | 12.50 | 11.75 | 13.00 | 12.38 | 0.88 | 7.10 | -1.00 | |
7 | 6.25 | 6.50 | 7.00 | 6.75 | 0.35 | 5.20 | 8.00 | |
25(OH)D3 | 1 | 500.00 | 490.25 | 528.50 | 509.38 | 27.05 | 5.30 | 1.90 |
2 | 250.00 | 241.25 | 243.00 | 242.13 | 1.24 | 0.50 | -3.20 | |
3 | 125.00 | 126.00 | 122.00 | 124.00 | 2.83 | 2.30 | -0.80 | |
4 | 62.50 | 60.00 | 62.75 | 61.38 | 1.94 | 3.20 | -1.80 | |
5 | 25.00 | 26.00 | 25.50 | 25.75 | 0.35 | 1.40 | 3.00 | |
6 | 12.50 | 11.75 | 13.75 | 12.75 | 1.41 | 11.10 | 2.00 | |
7 | 6.25 | 6.25 | 8.00 | 7.13 | 1.24 | 17.40 | 14.00 |
分析物 | 水平 | 理论值 | 测定值1 | 测定值2 | s | CV (%) | 偏倚(%) | |
---|---|---|---|---|---|---|---|---|
25(OH)D2 | 1 | 500.00 | 484.50 | 515.50 | 500.00 | 21.92 | 4.40 | 0.00 |
2 | 250.00 | 247.50 | 249.75 | 248.63 | 1.59 | 0.60 | -0.60 | |
3 | 125.00 | 129.50 | 124.00 | 126.75 | 3.89 | 3.10 | 1.40 | |
4 | 62.50 | 61.75 | 68.00 | 64.88 | 4.42 | 6.80 | 3.80 | |
5 | 25.00 | 22.00 | 21.75 | 21.88 | 0.18 | 0.80 | -12.50 | |
6 | 12.50 | 11.75 | 13.00 | 12.38 | 0.88 | 7.10 | -1.00 | |
7 | 6.25 | 6.50 | 7.00 | 6.75 | 0.35 | 5.20 | 8.00 | |
25(OH)D3 | 1 | 500.00 | 490.25 | 528.50 | 509.38 | 27.05 | 5.30 | 1.90 |
2 | 250.00 | 241.25 | 243.00 | 242.13 | 1.24 | 0.50 | -3.20 | |
3 | 125.00 | 126.00 | 122.00 | 124.00 | 2.83 | 2.30 | -0.80 | |
4 | 62.50 | 60.00 | 62.75 | 61.38 | 1.94 | 3.20 | -1.80 | |
5 | 25.00 | 26.00 | 25.50 | 25.75 | 0.35 | 1.40 | 3.00 | |
6 | 12.50 | 11.75 | 13.75 | 12.75 | 1.41 | 11.10 | 2.00 | |
7 | 6.25 | 6.25 | 8.00 | 7.13 | 1.24 | 17.40 | 14.00 |
项目 | 期望值 | 实测值1 | 实测值2 | 实测值3 | 实测值4 | 实测值5 | s | CV (%) | 偏倚(%) | S/N | |
---|---|---|---|---|---|---|---|---|---|---|---|
25(OH)D2 | 2.50 | 2.25 | 2.75 | 2.00 | 2.50 | 2.75 | 2.45 | 0.33 | 13.30 | -2.00 | 743∶1 |
1.25 | 1.00 | 1.50 | 2.00 | 1.00 | 0.75 | 1.25 | 0.50 | 40.00 | 0.00 | 494∶1 | |
0.63 | 1.00 | 1.00 | 0.50 | 0.75 | 0.50 | 0.75 | 0.25 | 33.33 | 20.00 | 325∶1 | |
25(OH)D3 | 2.50 | 2.75 | 2.75 | 3.25 | 3.25 | 2.50 | 2.90 | 0.34 | 11.57 | 16.00 | 141∶1 |
1.25 | 1.75 | 1.25 | 1.25 | 1.25 | 1.25 | 1.35 | 0.22 | 16.56 | 8.00 | 70∶1 | |
0.63 | 1.00 | 1.00 | 1.00 | 1.00 | 1.25 | 1.05 | 0.11 | 10.65 | 68.00 | 47∶1 |
项目 | 期望值 | 实测值1 | 实测值2 | 实测值3 | 实测值4 | 实测值5 | s | CV (%) | 偏倚(%) | S/N | |
---|---|---|---|---|---|---|---|---|---|---|---|
25(OH)D2 | 2.50 | 2.25 | 2.75 | 2.00 | 2.50 | 2.75 | 2.45 | 0.33 | 13.30 | -2.00 | 743∶1 |
1.25 | 1.00 | 1.50 | 2.00 | 1.00 | 0.75 | 1.25 | 0.50 | 40.00 | 0.00 | 494∶1 | |
0.63 | 1.00 | 1.00 | 0.50 | 0.75 | 0.50 | 0.75 | 0.25 | 33.33 | 20.00 | 325∶1 | |
25(OH)D3 | 2.50 | 2.75 | 2.75 | 3.25 | 3.25 | 2.50 | 2.90 | 0.34 | 11.57 | 16.00 | 141∶1 |
1.25 | 1.75 | 1.25 | 1.25 | 1.25 | 1.25 | 1.35 | 0.22 | 16.56 | 8.00 | 70∶1 | |
0.63 | 1.00 | 1.00 | 1.00 | 1.00 | 1.25 | 1.05 | 0.11 | 10.65 | 68.00 | 47∶1 |
样本 | 批内 | 批间 | ||||
---|---|---|---|---|---|---|
s(nmol/L) | CV(%) | s(nmol/L) | CV(%) | |||
样本1 | 13.22 | 0.51 | 3.84 | 12.97 | 0.67 | 5.20 |
样本2 | 119.02 | 3.72 | 3.12 | 119.32 | 6.18 | 5.18 |
样本3 | 334.40 | 7.74 | 2.31 | 330.15 | 11.02 | 3.34 |
样本 | 批内 | 批间 | ||||
---|---|---|---|---|---|---|
s(nmol/L) | CV(%) | s(nmol/L) | CV(%) | |||
样本1 | 13.22 | 0.51 | 3.84 | 12.97 | 0.67 | 5.20 |
样本2 | 119.02 | 3.72 | 3.12 | 119.32 | 6.18 | 5.18 |
样本3 | 334.40 | 7.74 | 2.31 | 330.15 | 11.02 | 3.34 |
样本浓度 | 加入量 | 理论浓度 | 实测值1 | 实测值2 | 实测值3 | 实测 | 回收率(%) | |
---|---|---|---|---|---|---|---|---|
25(OH)D2 | 25(OH)D3 | |||||||
31.1 | 2.84 | 8.52 | 42.44 | 41.50 | 37.25 | 40.00 | 39.58 | 93.26 |
8.52 | 2.84 | 42.44 | 44.75 | 44.00 | 43.00 | 43.92 | 103.49 | |
5.68 | 17.04 | 53.80 | 54.00 | 51.25 | 53.50 | 52.92 | 98.36 | |
17.04 | 5.68 | 53.80 | 56.25 | 55.50 | 52.25 | 54.67 | 101.62 | |
63.3 | 13.39 | 40.18 | 116.87 | 121.25 | 121.50 | 121.00 | 121.25 | 103.75 |
40.18 | 13.39 | 116.87 | 126.75 | 123.25 | 130.25 | 126.75 | 108.45 | |
26.79 | 80.36 | 170.44 | 161.50 | 173.00 | 186.50 | 173.67 | 101.90 | |
80.36 | 26.79 | 170.44 | 197.50 | 193.75 | 182.25 | 191.17 | 112.16 |
样本浓度 | 加入量 | 理论浓度 | 实测值1 | 实测值2 | 实测值3 | 实测 | 回收率(%) | |
---|---|---|---|---|---|---|---|---|
25(OH)D2 | 25(OH)D3 | |||||||
31.1 | 2.84 | 8.52 | 42.44 | 41.50 | 37.25 | 40.00 | 39.58 | 93.26 |
8.52 | 2.84 | 42.44 | 44.75 | 44.00 | 43.00 | 43.92 | 103.49 | |
5.68 | 17.04 | 53.80 | 54.00 | 51.25 | 53.50 | 52.92 | 98.36 | |
17.04 | 5.68 | 53.80 | 56.25 | 55.50 | 52.25 | 54.67 | 101.62 | |
63.3 | 13.39 | 40.18 | 116.87 | 121.25 | 121.50 | 121.00 | 121.25 | 103.75 |
40.18 | 13.39 | 116.87 | 126.75 | 123.25 | 130.25 | 126.75 | 108.45 | |
26.79 | 80.36 | 170.44 | 161.50 | 173.00 | 186.50 | 173.67 | 101.90 | |
80.36 | 26.79 | 170.44 | 197.50 | 193.75 | 182.25 | 191.17 | 112.16 |
样本编号 | 测定值1(nmol/L) | 测定值2(nmol/L) | 同方法组 | 偏倚(%) | |
---|---|---|---|---|---|
436 | 73.70 | 78.90 | 76.30 | 81.10 | -5.90 |
437 | 34.20 | 28.20 | 31.20 | 35.10 | -11.20 |
438 | 60.10 | 62.90 | 61.50 | 58.40 | 5.30 |
439 | 44.00 | 43.60 | 43.80 | 43.10 | 1.70 |
440 | 47.40 | 50.60 | 49.00 | 50.60 | -3.10 |
样本编号 | 测定值1(nmol/L) | 测定值2(nmol/L) | 同方法组 | 偏倚(%) | |
---|---|---|---|---|---|
436 | 73.70 | 78.90 | 76.30 | 81.10 | -5.90 |
437 | 34.20 | 28.20 | 31.20 | 35.10 | -11.20 |
438 | 60.10 | 62.90 | 61.50 | 58.40 | 5.30 |
439 | 44.00 | 43.60 | 43.80 | 43.10 | 1.70 |
440 | 47.40 | 50.60 | 49.00 | 50.60 | -3.10 |
[1] | WHITING SJ,LANGLOIS KA,VATANPARAST H,et al.The vitamin D status of Canadians relative to the 2011 Dietary Reference Intakes: an examination in children and adults with and without supplement use[J].Am J Clin Nutr, 2011,94(1):128-135. |
[2] | AUTIER P,GANDINI S.VITAMIN D supplementation and total mortality: a meta-analysis of randomized controlled trials[J]. Arch Intern Med,2007,167(16):1730-1737. |
[3] | PETERLIK M,GRANT WB,CROSS HS.Calcium,vitamin D and cancer[J]. Anticancer Res,2009,29(9):3687-3698. |
[4] | PEIRIS AN,BAILEY B,MANNING T,et al.Testing for vitamin D deficiency in veterans-is there a seasonal bias[J]. J Am Med Dir Assoc,2010,11(2):128-131. |
[5] | PIEPER CF,COLON-EMERIC C,CAMINIS J,et al.Distribution and correlates of serum 25-hydroxyvitamin D levels in a sample of patients with hip fracture[J]. Am J Geriatr Pharmacother,2007,5(4):335-340. |
[6] | SOUBERBIELLE JC,BODY JJ,LAPPE JM,et al.Vitamin D and musculoskeletal health,cardiovascular disease,autoimmunity and cancer: recommendations for clinical practice[J].Autoimmun Rev,2010,9(11):709-715. |
[7] | SOKOL SI,TSANG P,AGGARWAL V,et al.Vitamin D status and risk of cardiovascular events: lessons learned via systematic review and meta-analysis[J].Cardiol Rev,2011,19(4):192-201. |
[8] | HEIJBOER AC,BLANKENSTEIN MA,KEMA IP,et al.Accuracy of 6 routine 25-hydroxyvitamin D assays: influence of vitamin D binding protein concentration[J]. Clin Chem,2012,58(3):543-548. |
[9] | CAVALIER E,CARLISI A,BEKAERT AC,et al.Analytical evaluation of the new Abbott Architect 25-OH vitamin D assay[J]. Clin Biochem,2012,45(6):505-508. |
[10] | LEINO A,TURPEINEN U,KOSKINEN P.Automated measurement of 25-OH vitamin D3 on the Roche Modular E170 analyzer[J]. Clin Chem,2008,54(12):2059-2062. |
[11] | FARRELL CJ,MARTIN S,MCWHINNEY B,et al.State-of-the-art vitamin D assays: a comparison of automated immunoassays with liquid chromatography-tandem mass spectrometry methods[J].Clin Chem,2012,58(3):531-542. |
[12] | SNELLMAN G,MELHUS H,GEDEBORG R,et al.Determining vitamin D status: a comparison between commercially available assays[J]. PLoS One,2010,5(7):e11555. |
[13] | STEPMAN HC,VANDERROOST A,VAN UYTFANGHE K,et al.Candidate reference measurement procedures for serum 25-hydroxyvitamin D3 and 25-hydroxyvitamin D2 by using isotope-dilution liquid chromatography-tandem mass spectrometry[J]. Clin Chem,2011,57(3):441-448. |
[14] | TAI SS,BEDNER M,PHINNEY KW.Development of a candidate reference measurement procedure for the determination of 25-hydroxyvitamin D3 and 25-hydroxyvitamin D2 in human serum using isotope-dilution liquid chromatography-tandem mass spectrometry[J]. Anal Chem,2010,82(5):1942-1948. |
[15] | VOGESER M,KYRIATSOULIS A,HUBER E,et al.Candidate reference method for the quantification of circulating 25-hydroxyvitamin D3 by liquid chromatography-tandem mass spectrometry[J]. Clin Chem,2004,50(8):1415-1417. |
[16] | Food and Drug Administration. Guidance for industry bioanalytical method validation[R]. Rockville:FDA, 2001. |
[17] | CARTER GD.25-hydroxyvitamin D: a difficult analyte[J]. Clin Chem,2012,58(3):486-488. |
[18] | SHAH I, PETROCZI A,NAUGHTON DP.Method for simultaneous analysis of eight analogues of vitamin D using liquid chromatography tandem mass spectrometry[J]. Chem Cent J,2012,6(1):112. |
[19] | CLUSE ZN,FUDGE AN,WHITING MJ,et al.Evaluation of 25-hydroxy vitamin D assay on the immunodiagnostic systems iSYS analyser[J]. Ann Clin Biochem,2012, 49(Pt 2):159-165. |
[20] | CHEN Y,KINNEY L, BOŽOVIC A, et al. Performance evaluation of Siemens ADVIA Centaur and Roche MODULAR Analytics E170 total 25-OH vitamin D assays[J]. Clin Biochem, 2012, 45(16-17):1485-1490. |
[21] | HSU SA,SOLDO J, GUPTA M.Evaluation of two automated immunoassays for 25-OH vitamin D: comparison against LC-MS/MS[J]. J Steroid Biochem Mol Biol,2013,136:139-145. |
[22] | HERRMANN M,HARWOOD T,GASTON-PARRY O,et al.A new quantitative LC tandem mass spectrometry assay for serum 25-hydroxy vitamin D[J]. Steroids,2010,75(13-14):1106-1112. |
[1] | Expert Group on the Consensus on Androgen Testing in Polycystic Ovary Syndrome China Association for Promotion of Health Science and Technology-Fertility Protection and Preservation Committee. Consensus on androgen testing in polycystic ovary syndrome [J]. Laboratory Medicine, 2023, 38(3): 203-208. |
[2] | YANG Xiaodong, LI Quanle, PAN Qingqing, ZHOU Jing, XIE Yangmin, ZOU Jihua, SHEN Min, ZHANG Man. Establishment and performance evaluation of candidate reference measurement procedure for the determination of immunosuppressive drugs in human whole blood by ID-LC-MS/MS [J]. Laboratory Medicine, 2023, 38(3): 215-222. |
[3] | ZENG Yanfen, WU Quanming, ZHOU Huan, ZHANG Qiuqin, KANG Yanli, LI Yao, CHEN Xijun, HUANG Jiangang, CHEN Falin. Determination limits of a severe acute respiratory syndrome coronavirus 2 nucleic acid determination reagent in 4 determination systems [J]. Laboratory Medicine, 2023, 38(3): 261-266. |
[4] | HUANG Zhongqiang, DAI Di, HOU Tieying, LIN Faquan, LU Zhiming, LIU Xiaoqi, SHEN Minna, WU Jie, ZHANG Pingan, ZHOU Jun, XIAO Yanqun, WANG Hualiang. Multicenter performance evaluation of a 7-microRNA test kit [J]. Laboratory Medicine, 2023, 38(11): 1075-1081. |
[5] | LIU Yang, GENG Kunjing, LI Hongjiang, SHI Haoxi, CHEN Sisi. Relationship of 25(OH)D3 and liver inflammatory stage with the efficacy of antiviral therapy [J]. Laboratory Medicine, 2022, 37(7): 628-631. |
[6] | ZHANG Zhibin, YANG Wanyong, RUAN Fuwang, WEN Qinghui. Roles of total IgE,sIgE and 25(OH)D in the evaluation of infant atopic dermatitis [J]. Laboratory Medicine, 2022, 37(3): 217-220. |
[7] | XU Runhao, ZOU Chen, CAO Yun, CHEN Feng, LIU Yijing, ZHANG Shulin, ZHANG Jie. Roles of serum amino acid determinations in the diagnosis of early non-small-cell lung cancer [J]. Laboratory Medicine, 2022, 37(12): 1157-1163. |
[8] | ZHAO Xiaomin, JIANG Xiaoyu, SHAN Kaili, SHAN Tieying, LIU Xiuzhen. Roles of Flt-3L,TF and 25(OH)D in iron deficiency anemia among infants and children [J]. Laboratory Medicine, 2022, 37(11): 1039-1043. |
[9] | LI Dingfu, HUANG Jian, CAI Meiling, WANG Lizhen, JIAN Kaigan, YANG Feifei. Relationship between serum 25(OH)D and disease control in children with allergic rhinitis-asthma syndrome [J]. Laboratory Medicine, 2022, 37(1): 32-35. |
[10] | ZHAO Xueling, ZHENG Xin, XU Huamei, SHI Hongyao. Correlation between serum 25(OH)D,specific IgE and VAS score in children with allergic rhinitis [J]. Laboratory Medicine, 2021, 36(6): 615-617. |
[11] | LI Bingling, CHENG Yating, ZHAO Beibei, SHE Xuhui, DONG Heng, YU Mujun. Evaluation of surrogate matrices for serum 25-hydroxyvitamin D calibrators by LC-MS/MS [J]. Laboratory Medicine, 2021, 36(2): 213-218. |
[12] | FANG Huiling, GE Danhong, ZHU Yuqing, ZHU Lingfeng, ZHAO Xiaojun, CHEN Ziqi, WANG Zimeng. Analysis on the external quality assessment results for 25-hydroxyvitamin D of Shanghai from 2015 to 2019 [J]. Laboratory Medicine, 2021, 36(2): 219-224. |
[13] | XU Runhao, ZOU Chen, ZHANG Jie, LI Min, ZHANG Shulin. Application of serum bile acid spectrum in the differential diagnosis of pneumonia and lung cancer [J]. Laboratory Medicine, 2021, 36(1): 1-7. |
[14] | CHEN Gang, WANG Jing, DENG Qiang, CHEN Xi, HU Dong, ZHUANG Lidong. Correlation and risk assessment of serum 25(OH)D level and multiple thyroid function indicators in patients with thyroid cancer before operation [J]. Laboratory Medicine, 2020, 35(9): 885-888. |
[15] | ZOU Jihua, LI Quanle, SHEN Min, ZHANG Man, YANG Xiaodong, TU Minmin, FENG Dongfang, SUN Hongna. Establishment and performance evaluation of candidate reference measurement procedure for serum aldosterone by liquid chromatography-tandem mass spectrometry [J]. Laboratory Medicine, 2020, 35(8): 811-817. |
Viewed | ||||||
Full text |
|
|||||
Abstract |
|
|||||