检验医学 ›› 2020, Vol. 35 ›› Issue (6): 595-600.DOI: 10.3969/j.issn.1673-8640.2020.06.017

• 技术研究与评价·论著 • 上一篇    下一篇

新型血清CYFRA21-1检测电化学生物传感器的研制

王建富1, 李小平2(), 陈文虎3, 樊凯4, 钱苗苗4   

  1. 1.杭州市下城区朝晖街道社区卫生服务中心,浙江 杭州 310014
    2.浙江树人大学树兰国际医学院,浙江 杭州 310015
    3.浙江省肿瘤医院,浙江 杭州 310022
    4. 杭州电子科技大学,浙江 杭州 310018
  • 收稿日期:2019-10-15 出版日期:2020-06-30 发布日期:2020-07-01
  • 作者简介:null

    作者简介:王建富,男,1962年生,副主任技师,主要从事临床医学检验工作。

  • 基金资助:
    浙江省科技厅科研资助项目(2017C37020)

A novel electrochemical biosensor for the detection of serum CYFRA21-1

WANG Jianfu1, LI Xiaoping2(), CHEN Wenhu3, FAN Kai4, QIAN Miaomiao4   

  1. 1. Community Health Service Center of Zhaohui Street of Xiacheng District,Hangzhou 310014,Zhejiang,China
    2. Shulan International Medical College,Zhejiang Shuren University,Hangzhou 310015,Zhejiang,China
    3. Zhejiang Cancer Hospital,Hangzhou 310022,Zhejiang,China
    4. Hangzhou Dianzi University,Hangzhou 310018,Zhejiang,China
  • Received:2019-10-15 Online:2020-06-30 Published:2020-07-01

摘要:

目的 构建一种用于检测细胞角蛋白19片段(CYFRA21-1)的高灵敏电化学生物传感器。方法 合成三维石墨烯@金纳米粒子(3D-G@Au)复合材料,用其修饰玻碳电极,通过壳聚糖、戊二醛与抗CYFRA21-1抗体交联,构建用于检测CYFRA21-1的3D-G@Au的电化学生物传感器。对构建成功的3D-G@Au的电化学生物传感器的电化学性能及检测性能进行初步评价。结果 循环伏安图显示裸玻碳电极、3D-G修饰玻碳电极和3D-G@Au修饰玻碳电极的氧化还原峰逐渐升高,提示3D-G@Au电化学生物传感器构建成功。传感器抗原-抗体相互作用的最佳温育时间为60 min,最佳孵育温度为35 ℃。采用3D-G@Au电化学生物传感器检测CYFRA21-1,线性范围为0.1~300.0 ng/mL,检测限为0.1 ng/mL(信噪比=3),10 mmol/L 维生素C、10 mmol/L 多巴胺和10 mmol/L 尿酸对检测无干扰,重现性较好[相对标准偏差(RSD)分别为1.08%、3.11%],与化学发光法比较,RSD为1.96%~4.78%。制备好的3D-G@Au电化学生物传感器可在4 ℃条件下稳定30 d。结论 构建的3D-G@Au电化学生物传感器可用于血清CYFRA21-1的检测。

关键词: 细胞角蛋白19片段, 三维石墨烯, 金纳米粒子, 电化学生物传感器

Abstract:

Objective To construct a highly sensitive electrochemical biosensor for the detection of cytokeratin 19-fragment(CYFRA21-1). Methods Three-dimensional graphene @gold nanoparticles(3D-G@Au)nanocomposite was modified on glassy carbon electrode surface to enhance the conductivity of electrochemical biosensor. Anti-CYFRA21-1 was captured and fixed on the modified 3D-G@Au glassy carbon electrode through the cross-linking of chitosan and glutaraldehyde. The 3D-G@Au electrochemical biosensor had been completed as to determine CYFRA21-1. The electrochemical and determination performance of successfully constructed 3D-G@Au electrochemical biosensor was evaluated preliminarily. Results The cyclic voltammetry showed that the redox peaks of bare glassy carbon electrode,3D-G modified glassy carbon electrode and 3D-G@Au modified glassy carbon electrode increased gradually,indicating that the 3D-G@Au electrochemical biosensor was successfully constructed. The optimal incubation time of antigen-antibody interaction was 60 min,and the optimal incubation temperature was 35 ℃. CYFRA21-1 was detected by 3D-G@Au electrochemical biosensor,the linear range was 0.1-300.0 ng/mL,and the determination limit was 0.1 ng/mL(signal-to-noise ratio=3). 10 mmol/L Vit C,10 mmol/L DA and 10 mmol/L UA had no interference to the determination,and the reproducibility was good [relative standard deviations(RSD) were 1.08% and 3.11%], compared with that of chemiluminescence method (RSD was 19.6%-4.78%). The 3D-G@Au electrochemical biosensor can be stable for 30 d at 4 ℃. Conclusions It is potential for the determination of serum CYFRA21-1 with 3D-G@Au electrochemical biosensor.

Key words: Cytokeratin 19-fragment, Three-dimensional graphene, Gold nanoparticle, Electrochemical biosensor

中图分类号: