The accuracy evaluation of nucleated red blood cell count determined by SYSMEX XN-1000 automatic hematology analyzer

WANG Yefei,ZHOU Yi,DING Lei,XIA Wenquan,HU Yiqun.   

  1. Faculty of Medical Laboratory Science,Ruijin Hospital,Shanghai Jiaotong University School of Medicine,Shanghai 200025,China
  • Received:2013-08-12 Online:2014-03-30 Published:2014-04-19

Abstract:

Objective To investigate the accuracy of nucleated red blood cell(NRBC) count determined by SYSMEX XN-1000 automatic hematology analyzer(XN-1000). Methods A total of 66 peripheral blood samples were collected and determined for NRBC by XN-1000,and the results of NRBC% were compared with those by flow cytometry (FCM) and manual microscopy for the same sample. The manual microscopy was as golden standard. Results There was a good correlation for NRBC between XN-1000 and microscopy(r=0.927),but the correlation was lower than that between FCM and manual microscopy (r=0.956).The correlation between XN-1000 and FCM was the best (r=0.994). The results of XN-1000 and FCM with manual microscopy had no statistical significance(P>0.05). XN-1000 for NRBC% had a better specificity (95.7%) than FCM. Conclusions XN-1000 automatic hematology analyzer has good performance for detecting NRBC in peripheral blood rapidly,accurately and effectively.

Key words: Sysmex XN-1000 automatic hematology analyzer, Nucleated red blood cell, Accuracy