›› 2012, Vol. 27 ›› Issue (8): 667-670.

Previous Articles     Next Articles

Evaluation on nucleated red blood cell detection by XE-5000 automated hematology analyzer

  

  1. Blood Disease Hospital,Institute of Hematology,Chinese Academy of Medical Science and Peking Union Medical College, Tianjin 300020,China
  • Received:2012-05-15 Online:2012-08-30 Published:2012-08-08

Abstract: Objective To evaluate the clinical application significance and methodology characteristic of Sysmex XE-5000 automated hematology analyzer (XE-5000) to count nucleated red blood cell(NRBC).  Methods The repeatability,linear range and carryover rate of XE-5000 were evaluated. A total of 391 blood samples were measured, and the alarm and quantitative determination of XE-5000 for NRBC were evaluated. The results were compared with manual microscopy. Results The XE-5000 results showed good repeatability, linear range and low carryover rate. When NRBC%>5.0%,it caused false increasing of leucocyte and lymphocyte. The results of manual microscopy were as a gold standard. The correlation was good between the XE-5000 and manual microscopy(r=0.997). The sensitivity for NRBC was 100.0%,and the specificity was 85.4%.  Conclusions NRBC counting by XE-5000 is sensitive,accurate and reliable,which is suitable for the clinical detection of NRBC in peripheral blood.

Key words: Nucleated red blood cell, Hematology analyzer, Evaluation