Laboratory Medicine ›› 2017, Vol. 32 ›› Issue (3): 214-218.DOI: 10.3969/j.issn.1673-8640.2017.03.014

• Orginal Article • Previous Articles     Next Articles

Precision and comparability assessment of Mindray H50 glycated hemoglobin A1c analyzer

GUO Zonghui1, CHEN Yongqiang1, ZENG Fangyin1   

  1. 1. Department of Clinical Laboratory,Peking University Care Luzhong Hospital,Zibo 255400,Shandong,China
    2. Department of Clinical Laboratory,the Fifth Affiliated Hospital of Southern Medical University,Guangzhou 510515,Guangdong,China
  • Received:2016-06-03 Online:2017-03-30 Published:2017-04-11

Abstract:

Objective To evaluate the precision and comparability of Mindray H50 glycated hemoglobin A1c (HbA1c)analyzer. Methods According to the Clinical and Laboratory Standards Institute(CLSI) EP5-A2 and EP9-A2,the precision and comparability of H50 were evaluated and compared with TOSOH HLC-723 G8 analyzer. Results The within-run,between-run,inter-day and total precisions [coefficients of variation(CV)] for clinical samples with high and medium values of HbA1c were 0.34%,0.77%,0.60%,1.03% and 0.86%,1.55%,0.00%,1.72%,respectively. There was a good comparability between H50 and G8(YH50=1.053 3XG8-0.486 7,r=0.998 4). The biases [95% confidence interval(CI)] were -0.27%(-0.30%--0.25%),-0.17%(-0.18%--0.15%),-0.14%(-0.16%--0.12%),-0.11%(-0.13%--0.10%),-0.01%(-0.02%-0.01%)at different HbA1c levels of 4.0%,6.0%,6.5%,7.0% and 9.0%. Conclusions H50 has a good precision and shows a correlation with G8,and it meets the requirements of the diagnosis and glucose monitoring for diabetes mellitus.

Key words: Glycated hemoglobin A1c, Diabetes mellitus, Precision, Comparability