检验医学 ›› 2015, Vol. 30 ›› Issue (2): 108-112.DOI: 10.3969/j.issn.1673-8640.2015.02.003

• 质谱技术在临床微生物检验中的应用专题 • 上一篇    下一篇

MALDI-TOF MS在中段尿样本细菌直接检测中的应用

王琳, 李佳萍, 陈功祥, 张嵘   

  1. 浙江大学医学院附属第二医院检验科,浙江 杭州 310009
  • 收稿日期:2014-10-28 出版日期:2015-02-28 发布日期:2015-02-12
  • 作者简介:null

    作者简介:王琳,男,1981年生,学士,主管技师,主要从事微生物检验工作。

    通讯作者:陈功祥,联系电话:0571-87783646。

Application of MALDI-TOF MS for identifying bacteria directly in midstream urine samples

WANG Lin, LI Jiaping, CHEN Gongxiang, ZHANG Rong   

  1. Department of Laboratory Medicine, the Second Affiliated Hospital of Zhejiang University School of Medicine, Zhejiang Hangzhou 310009
  • Received:2014-10-28 Online:2015-02-28 Published:2015-02-12

摘要: 目的

运用基质辅助激光解析电离飞行时间质谱(MALDI-TOF MS)鉴定系统对中段尿样本中的细菌进行直接快速检测,以建立快速的细菌检测方法便于临床应用。

方法

692例中段尿样本接种于固体培养基上,利用MALDI-TOF MS和Vitek 2 Compact全自动微生物分析系统(简称Vitek 2 Compact)对单个菌落进行鉴定;同时对尿液样本进行前处理后直接进行质谱分析。

结果

692例中段尿样本中,98例单一菌感染且细菌数>104 CFU/mL的样本应用MALDI-TOF MS直接检测出74株菌,检出率达75.5%,其中大肠埃希菌39株,检出率达79.6%;粪肠球菌23株,检出率为100.0%;摩根摩根菌5株,检出率为100.0%。64例样本为2种菌混合感染,采用MALDI-TOF MS直接检测出44例样本中的1种优势细菌。质谱直接检测尿液的鉴定结果与单菌落质谱鉴定结果及Vitek 2 Compact鉴定结果完全一致,符合率为100.0%。

结论

MALDI-TOF MS可用于直接检测尿液样本中的细菌,检出率高,结果可靠;MALDI-TOF MS对混合菌感染的样本有一定鉴定能力;该方法可比常规尿培养鉴定方法提早42 h报告鉴定结果,是对临床尿液样本细菌鉴定的一大革命性进步。

关键词: 基质辅助激光解析电离飞行时间质谱, 中段尿, 细菌直接检测

Abstract: Objective

To construct rapid identification method for clinical laboratory with the use of matrix-assisted laser desorption/ionization-time of flight mass spectrometry (MALDI-TOF MS) for the direct identification of bacteria in midstream urine samples.

Methods

A total of 692 midstream urine samples were cultivated on agar,and were identified by MALDI-TOF MS and Vitek 2 Compact automated microbial amalyslis system (Vitek 2 Compact). Direct identification of midstream urine samples were also conducted by MALDI-TOF MS after preparation.

Results

Of the 692 midstream urine samples, 98 samples were positive for one kind of bacteria (>104 CFU/mL). The overall detection rate of direct identification by MALDI-TOF MS was 75.5%,and 74 isolates were identified,among which 39 isolates were Escherichia coli with the detection rate of 79.6%,23 isolates were Enterococcus faecalis with the detection rate of 100.0%,and 5 isolates were Morganella morganii with the detection rate of 100.0%. Among the 64 midstream urine samples with multiple microorganisms,44 samples were detected with a dominant bacteria by direct identification. The consistency rate was 100.0% among the results by direct identification,the identification of colony by MALDI-TOF MS and Vitek 2 Compact.

Conclusions

MALDI-TOF MS could be applied to identify directly bacteria in midstream urine samples with high detection rate and accuracy, and MALDI-TOF MS also shows certain performance in identifying samples containing multiple bacteria. The identification results could be 42 h earlier available than regular methods. It is a revolutionary improvement for the clinical identification of bacteria in urine samples.

Key words: Matrix-assisted laser desorption/ionization-time of flight mass spectrometry, Midstream urine, Bacterial direct identification

中图分类号: