Laboratory Medicine ›› 2017, Vol. 32 ›› Issue (6): 524-530.DOI: 10.3969/j.issn.1673-8640.2017.06.018

• Orginal Article • Previous Articles     Next Articles

Establishment and performance evaluation of liquid-chromatography tandem-mass spectrometry for determining serum cortisone

ZHU Yuqing, ZHU Lingfeng, ZHAO Xiaojun, XU Chong, SHAO Weijie, CAO Danru, LU Yinhua   

  1. Department of Clinical Immunology,Shanghai Center for Clinical Laboratory,Shanghai 200126,China
  • Received:2016-11-25 Online:2017-06-30 Published:2017-07-04

Abstract:

Objective To establish isotope-dilution liquid-chromatography tandem-mass spectrometry(ID-LC-MS/MS) for determining serum cortisone,and to evaluate its performance. Methods ABSCIEX TRIPLE QUAD 5500 liquid-chromatography tandem-mass spectrometry was used to determine serum cortisone quantitatively. According to the Clinical and Laboratory Standards Institute(CLSI) document,the performance,including linearity,precision,extraction and recycling rates and accuracy,was verified. Results Cortisone-d4 was added to analytes, and they were equilibrated at room temperature for 1 h. ID-LC-MS/MS for determining serum cortisone had linear regression equation Y=0.018 7X+0.043 4 (r=0.999 8). The linear range was 25-500 ng/mL. The determination limit was 0.05 ng/mL for serum cortisone. The within-run and between-run coefficients of variation(CV) were <4% and <6%. The extraction and recycling rates were 84.5%-90.4%. The biases of the International Federation of Clinical Chemistry and Laboratory Medicine External Quality Assessment Scheme for Reference Laboratories in Laboratory Medicine(RELA)-A and RELA-B were 3.19% and 2.84%. Conclusions The established ID-LC-MS/MS has good precision,accuracy and repeatability,which is suitable for determining serum cortisone.

Key words: Isotope-dilution liquid-chromatography tandem-mass spectrometry, Performance verification, Cortisone

CLC Number: