Performance verification of CS-5100 automatic blood coagulation analyzer for the determinations of antithrombin Ⅲ,protein C and protein S activities

LI Wen, HE Aili, ZHANG Wen, ZHOU Ting   

  1. Department of Clinical Laboratory,the Second Affiliated Hospital of Xi'an Jiaotong University,Xi'an 710004,Shaanxi,China
  • Received:2017-08-23 Online:2018-11-30 Published:2018-11-28

Abstract:

Objective To evaluate the performance of CS-5100 automatic blood coagulation analyzer for the determinations of antithrombin Ⅲ(AT Ⅲ),protein C(Pc) and protein S(Ps) activities. Methods According to the Clinical and Laboratory Standards Institute (CLSI) guidelines,the imprecision [coefficient of variation (CV)],accuracy,linearity,carry-over rate and reference interval of CS-5100 for the determinations of AT Ⅲ,Pc and Ps activities were validated. Results The within-run CV of CS-5100 for the determinations of AT Ⅲ,Pc and Ps activities were <1.9%,and the inter-day CV were <4.1%,meeting manufacturers' requirements. The bias of accuracy verification results was within biology CV requirements. The linear regression equation between AT Ⅲ linear validation theoretical value and measured values met the slope of 1±0.05,the determination coefficient (r2)≥0.95 and carry-over CV<10%. Reference interval verification results showed that the determination index R was ≥0.9. Conclusions The precision,accuracy,linearity,carry-over rate and reference interval of CS-5100 meet the requirements of quality control,which can ensure determination quality.

Key words: Automatic blood coagulation analyzer, Antithrombin Ⅲ, Protein C, Protein S, Performance verification

CLC Number: