›› 2013, Vol. 28 ›› Issue (4): 312-314.DOI: 10.3969/j.issn.1673-8640.2013.04.015

Previous Articles     Next Articles

The investigation of the retesting range of the TPAb sample detected by ELISA

  

  1. Funing People′s Hospital,Jiangsu Funing 224400, China
  • Received:2012-04-09 Revised:2012-05-08 Online:2013-04-30 Published:2013-04-05

Abstract: Objective To investigate the retesting range of anti-treponema pallidum antibody (TPAb) sample detected by enzyme-linked immunosorbent assay(ELISA).  Methods A total of 1 257 samples were collected in the first TPAb testing with absorbance (A) value > 0.08 by ELISA. They were retested duplicately by the same reagent kit of same lot number. The samples retested with A vaule > 0.14 were comfirmed by treponema pallidum particale agglutination (TPPA) test.The appropriate retesting range was analyzed statistically. Results The samples were classified into 7 groups according to the first A value by ELISA: 0.08-0.14,0.15-0.30,0.31-0.60,0.61-0.90,0.91-1.20,1.21-2.00 and >2.00. The retesting coincidence rates by ELISA were 16.6%(37/223),21.6%(56/259),32.0%(66/206),50.5%(100/198),78.0%(99/127),94.8%(147/155) and 100.0%(89/89). The coincidence rates by TPPA and retesting by ELISA were 0.0%(0/37),3.6%(2/56),22.7%(15/66),38.0%(38/100),80.8%(80/99),93.2%(137/147) and 96.6%(86/89). Conclusions The A value of the retesting range by ELISA is considered appropriate from 0.14-1.20 under a standardized laboratory condition.The appropriate retesting range can assure the accuracy of the testing results and can avoid the waste of manpower and reagents.

Key words: Retesting range, Anti-treponema pallidum antibody, Enzyme-linked immunosorbent assay, Treponema pallidum particale agglutination test