Laboratory Medicine ›› 2019, Vol. 34 ›› Issue (3): 253-258.DOI: 10.3969/j.issn.1673-8640.2019.03.015

Previous Articles     Next Articles

Comparison of the results' consistency of 3 automated coagulation analyzers for coagulation items

WANG Yue1, QU Chenxue2, GONG Yan2, GAO Xueshuo1, WANG Guohong1, XU Guobin1   

  1. 1. Department of Clinical Laboratory,Key Laboratory of Carcinogenesis and Translational Research,Ministry of Education,Peking University Cancer Hospital and Institute,Beijing 100142,China
    2. Department of Clinical Laboratory,the First Hospital of Peking University,Beijing 100034,China
  • Received:2018-08-01 Online:2019-03-30 Published:2019-03-27

Abstract:

Objective To investigate the consistency of 3 automated coagulation analyzers for coagulation items [prothrombin time(PT),international normalized ratio(INR),activated partial thromboplastin time(APTT),fibrinogen(Fib),thrombin time(TT),D-dimer(DD)and fibrin/fibrinogen degradation product(FDP)] and the matrix effects of quality control materials. Methods A total of 100 samples were collected. Totally,PT,PT-INR,APTT,Fib,TT,DD and FDP were determined by 3 automated coagulation analyzers,including CA7000,Top700 and CP2000. Linear regression analysis was used to analyze the correlation of results. The relative deviation between any 2 analyzers was calculated. The matrix effect of 18 quality control materials for PT,Fib and DD was assessed with regression. Results There were good correlations among the 3 automated coagulation analyzers for PT,PT-INR and Fib (r2=0.91-0.99). The deviations between any 2 analyzers were <15%,15% and 20%. The results of DD among the 3 automated coagulation analyzers had correlations (r2≥0.90),CA7000 and CP2000 had good consistency,but the results of Top700 were lower than those of CA7000 and CP2000. Moderate correlations were observed for APTT and TT(r2=0.56-0.76). Medical kits for FDP and matched calibrators were adopted to the 3 analyzers. There was a high correlation for FDP and DD (r2>0.99,r2=0.92-0.99),and there was difference between any 2 analyzers. There was no compatibility for PT,Fib and DD on the 3 analyzers in 18 quality control materials. Conclusions The consistency of the results of the 3 automated coagulation analyzers is diverse. The matrix effects should be considered,which makes the deviation of quality control materials greater than that of patient.

Key words: Coagulation test, Correlation, Quality control material, Matrix effect