Laboratory Medicine ›› 2018, Vol. 33 ›› Issue (10): 928-929.DOI: 10.3969/j.issn.1673-8640.2018.10.013

Previous Articles     Next Articles

Comparison on peripheral blood and venous blood for the determination results of MQ-6000 glycated hemoglobin analyzer

QIAN Cheng, MA Jing, XUE Mingyang, ZHAI Feng, LIU Tong   

  1. Department of Endocrinology,Renji Hospital,Shanghai Jiaotong University School of Medicine,Shanghai 200127,China
  • Received:2018-05-16 Online:2018-10-30 Published:2018-10-23

Abstract:

Objective To investigate the comparability of glycated hemoglobin A1c(HbA1c) determination using finger peripheral blood and venous blood by MQ-6000 glycated hemoglobin analyzer(MQ-6000) and the clinical application value of peripheral blood for the determination of HbA1c. Methods A total of 97 patients with diabetes mellitus were enrolled,and 39 healthy subjects were enrolled as control group. Finger peripheral blood and venous blood were used for HbA1c determination by MQ-6000,and the results were compared with those using venous blood determined by VARIANT Ⅱ hemoglobin testing system(VARIANT Ⅱ). Results There was no statistical significance for the results of HbA1c determination between using peripheral blood and venous blood in control and diabetes mellitus groups(P>0.05),and there was a good correlation between using peripheral blood and venous blood(r=0.996). MQ-6000 had a good correlation with VARIANT Ⅱ(r=0.992). Conclusions Finger peripheral blood can replace venous blood for the determination of HbA1c and be used for the screening,diagnosis and monitoring of diabetes mellitus.

Key words: Glycated hemoglobin A1c, Peripheral blood, Venous blood, Diabetes mellitus