Establishment and performance of nano particle-enhanced turbidimetric immunoassay for determining serum lactate dehydrogenase isoenzyme 3

DONG Yiyu1, WU Yizheng2, SHEN Lili3, CHEN Qun3   

  1. 1. Department of Clinical Laboratory,Rongjun Hospital of Zhejiang,Jiaxing 314000,Zhejiang,China
    2. Department of Clinical Medicine,Zhejiang University School of Medicine,Hangzhou 310029,Zhejiang,China
    3. Department of Internal Medicine,Rongjun Hospital of Zhejiang,Jiaxing 314000,Zhejiang,China
  • Received:2016-02-03 Online:2017-03-30 Published:2017-03-30

Abstract:

Objective To establish nano particle-enhanced turbidimetric immunoassay(PETIA) for determining serum lactate dehydrogenase isoenzyme 3(LDH3).Methods The optimal reaction conditions [antibody concentration,latex particle size,1-ethyl-3-(3-dimethylaminopropyl)carbodiimide(EDAC)/ N-hydroxysuccinimido(NHS) concentration,the kind and pH value of buffer solution and reaction time] were determined. PETIA for determining serum LDH3 was established. According to the National Committee for Clinical Laboratory Standards EP5-A2,the precision,linear range,recovery rate,anti-interference ability and analytic sensitivity were evaluated,and the initial reference interval was established.Results The optimal reaction conditions were antibody concentration 0.1 mg/mL,latex particle size 120 nm,EDAC/ NHS concentration 10 mg/mL,phosphate buffered saline(PBS) with pH value 7.8 and coupling rate remaining unchanged under room temperature(25 ℃)and low temperature(4 ℃) for 24 h. For LDH3 with low value(16.5 U/L),median value(65.0 U/L) and high value(155.0 U/L),the within-run coefficients of variation(CV) were 5.77%,4.02% and 3.77%,and the between-run CV were 6.59%,4.44% and 3.93%. The linear range was 3.6-200.0 U/L. The average recovery rate was 102.1%. The factors of 900 U/L rheumatoid factor,500 μmol/L bilirubin,9 mmol/L triglyceride,5 g/L hemoglobin,5 g/L vitamin C had no interference for PETIA in the determination of LDH3. The analytic sensitivity was 3.6 U/L. The reference interval was 13.5-75.9 U/L. Conclusions PETIA for the determination of LDH3 is simple,rapid,sensitive and accurate,and it can be used in automatic biochemical analyzers,which can meet the requirements of clinical determination.

Key words: Lactate dehydrogenase isoenzyme 3, Nano particle-enhanced turbidimetric immunoassay, Performance evaluation