Evaluation on the clinical application of Sysmex UF-1000i urine formed element analyzer for screening funguria

  

  1. Department of Clinical Laboratory,Changzhou Traditional Chinese Medicine Hospital,Nanjing Traditional Chinese Medicine University,Jiangsu Changzhou 213003,China
  • Received:2012-05-15 Revised:2012-07-09 Online:2013-04-30 Published:2013-04-05

Abstract: Objective To evaluate the clinical application significance of Sysmex UF-1000i urine formed element analyzer (UF-1000i analyzer) for screening funguria.   Methods The precision,linear range and carryover rate of Candida albicans were detected by UF-1000i analyzer. A total of 200 urine samples which UF-1000i analyzer showed like yeast cell(YLC) positive and 400 YLC negative urine samples were detected by manual microscopy. Results For UF-1000i analyzer detecting Candida albicans,the precision was 3.83%- 4.71%,the linear range was 10.0-5 873.1 cells/μL,the correlation coefficient was 0.999, and the carryover rate was 0.12%. Manual microscopy was as the gold standard,and the result of receiver operating characteristic (ROC) curve showed the best cut-off value for UF-1000i analyzer detecting YLC was 18.05 cells/μL. The sensitivity,specificity,accuracy,positive predictive value and negative predictive value of UF-1000i analyzer in the diagnosis of funguria were 97.0%,85.6%,87.5%,57.4% and 99.3%,respectively. Conclusions UF-1000i analyzer can provide quantitative results of fungal spores. It aids to diagnose urinary tract fungal infection.

Key words: Funguria, Sysmex UF-1000i urine formed element analyzer, Microscopy