Laboratory Medicine ›› 2018, Vol. 33 ›› Issue (2): 152-155.DOI: 10.3969/j.issn.1673-8640.2018.02.012

• Orginal Article • Previous Articles     Next Articles

Application of AVE-562 automated feces analyzer

BAO Yunwen1, GU Kongzhen2, XU Keqian2   

  1. 1. Department of Clinical Laboratory,Sun Yat-Sen Memorial Hospital,Sun Yat-Sen University,Guangzhou 510120,Guangdong,China
    2. Department of Laboratory Medicine,Xiangya School of Medicine,Central South University,Changsha 410013,Hunan,China;
  • Received:2016-11-08 Online:2018-02-28 Published:2018-03-02

Abstract:

Objective To evaluate the application of AVE-562 automated feces analyzer. Methods A total of 3 028 feces specimens were determined by manual microscopy and AVE-562 automated feces analyzer. The results of manual microscopy were as gold standard. Occult blood(OB) test results determined automatically by AVE-562 were compared with those of manual microscopy by OB test strip of Kangzhu Company. The sensitivity,specificity and accuracy were calculated. Results The coincidence rate was 97.85% for OB test of AVE-562 and OB test strip(kappa=0.918). The sensitivities of red blood cell(RBC),white blood cell(WBC),parasite and fat globule were 86.05%,80.95%,72.00% and 91.89%,respectively. The specificities of RBC,WBC,parasite and fat globule were 99.63%,99.80%,99.73% and 99.86%,respectively. The accuracies of RBC,WBC,parasite and fat globule were 99.44%,99.54%,99.50% and 99.67%,respectively. The kappa values of RBC,WBC,parasite and fat globule were 0.810,0.827,0.703 and 0.930,respectively. Conclusions The results of RBC,WBC and other indices by AVE-562 automated feces analyzer and manual microscopy have good consistency. It is easy to operate for AVE-562 automated feces analyzer with application value.

Key words: Feces examination, Feces analyzer, Manual microscopy

CLC Number: