2种国产化学发光检测系统与电化学发光检测系统检测CA72-4的一致性评价

魏衍财, 郑维玲, 石燕, 宋妙丽, 叶科学, 卢旬, 杨辰   

  1. 南京医科大学附属苏州医院医学检验科,江苏 苏州 215002
  • 收稿日期:2017-10-27 出版日期:2018-09-30 发布日期:2018-10-13
  • 作者简介:null

    作者简介:魏衍财,男,1986年生,硕士,主管技师,主要从事临床标记免疫检测工作。

Consistency of electrochemiluminescence and 2 domestic chemiluminescence determination systems in CA72-4 determination

WEI Yancai, ZHENG Weiling, SHI Yan, SONG Miaoli, YE Kexue, LU Xun, YANG Chen   

  1. Department of Clinical Laboratory,Suzhou Hospital,Nanjing Medical University,Suzhou 215002,Jiangsu,China
  • Received:2017-10-27 Online:2018-09-30 Published:2018-10-13

摘要:

目的 评价电化学发光检测系统和2种国产化学发光检测系统检测糖类抗原72-4(CA72-4)结果的一致性。方法 根据美国临床实验室标准化协会(CLSI)EP09-A3文件,以cobas e601电化学发光免疫分析仪及配套试剂组成的检测系统(简称e601检测系统)作为参比系统(X),以MAGLUMI 4000全自动化学发光分析仪及配套试剂组成的检测系统(简称MAGLUMI 4000检测系统)作为待评系统1(Y1),以CL-2000i化学发光分析仪及配套试剂组成的检测系统(CL-2000i检测系统)为待评系统2(Y2)。收集42例覆盖线性范围的临床血清样本,每份样本用3种检测系统进行单次检测。使用广义极端学生化偏差(ESD)法进行离群值检验,选择最优的回归模型拟合回归方程,计算医学决定水平处的偏移,以1/2室间质评平均偏移(±7.65%)作为可接受标准。结果 散点图显示e601检测系统(X)与MAGLUMI 4000检测系统(Y1)和CL-2000i检测系统(Y2)的相关性均良好,目测各发现1个离群值,广义ESD法确认CA72-4检测结果无离群值。通过Deming回归分析(XY1)和Passing-Bablok回归分析(XY2)得回归方程分别为Y1=5.020 6+1.149 8X,Y2=0.726 3+0.830 0X。将CA72-4的医学决定水平(6.9 U/mL)代入回归方程,计算得e601检测系统(X)和MAGLUMI 4000检测系统(Y1)的相对偏移为60.96%,其相对偏移大于1/2室间质评偏移,比对不可接受;e601检测系统(X)与CL-2000i检测系统(Y2)的相对偏移为-6.74%,比对结果可以接受。结论 e601检测系统与MAGLUMI 4000检测系统之间CA72-4的检测结果不具有可比性,e601检测系统与CL-2000i检测系统之间CA72-4的检测结果具有可比性。CLSI EP09-A3文件可用于不同标记免疫检测系统之间的可比性研究。

关键词: 糖类抗原72-4, 化学发光免疫分析, 电化学发光免疫分析, 方法学比对

Abstract:

Objective To evaluate the consistency of electrochemiluminescence and 2 domestic chemiluminescence determination systems in carbohydrate antigen 72-4(CA72-4)determination. Methods According to the Clinical and Laboratory Standards Institute(CLSI)EP09-A3 document,42 clinical serum specimens in linear range were determined by cobas e601 electrochemiluminescence immunoassay system(reference method,X),MAGLUMI 4000 chemiluminescence immunoassay system(experimental method 1,Y1),CL-2000i chemiluminescence immunoassay system(experimental method 2,Y2). The results of outliers were determined by generalized extreme studentized deviate(ESD)method. The optimal regression model was fitted to the regression equation. The biases were calculated at the medical decision level,and the comparability were judged by 1/2 external quality assessment bias(±7.65%) as acceptance standard. Results The correlations between e601(X) and MAGLUMI 4000(Y1) and between e601(X) and CL-2000i(Y2)were good by plotting scatter plot. One outlier was visually found in each group. No outlier of CA72-4 was found using generalized ESD method. Deming(X and Y1)and Passing-Bablok(X and Y2)were used. The regression equations were Y1 = 5.020 6+1.149 8X and Y2= 0.726 3+0.830 0X,respectively. Assuming CA72-4 medical decision level of 6.9 U/mL to regression equation,the proportional bias was 60.96% at e601(X) and MAGLUMI 4000(Y1),which was higher than acceptable value(1/2 bias). The proportional bias was -6.74% at e601(X) and CL-2000i(Y2),which was lower than acceptable value(1/2 bias). Conclusions The results of CA72-4 determination by e601 and MAGLUMI 4000 are not comparable,while results of CA72-4 determination by e601 and CL-2000i are comparable. CLSI EP09-A3 document can be used for comparison between different labeled immunoassay determination systems.

Key words: Carbohydrate antigen 72-4, Chemiluminescence immunoassay, Electrochemiluminescence, Methodology comparison