检验医学 ›› 2018, Vol. 33 ›› Issue (8): 679-682.DOI: 10.3969/j.issn.1673-8640.2018.08.003

• • 上一篇    下一篇

基于液相色谱串联质谱技术对血清同型半胱氨酸常规检测系统的正确度评价

钟巧玲1, 孙骥2, 翁旭琦3, 冯萍3, 谢丹芸3   

  1. 1. 重庆市巴南区第二人民医院检验科,重庆 400054
    2.宁波出入境检验检疫局技术中心,浙江 宁波 315000
    3. 美康生物参考实验室,浙江 宁波 315104
  • 收稿日期:2018-03-20 出版日期:2018-08-10 发布日期:2018-09-07
  • 作者简介:null

    作者简介:钟巧玲,女,1976年生,学士,副主任技师,主要从事临床免疫检验工作。

Accuracy evaluation of routine measurement procedure for serum homocysteine based on liquid chromatography-tandem mass spectrometry

ZHONG Qiaoling1, SUN Ji2, WENG Xuqi3, FENG Ping3, XIE Danyun3   

  1. 1. Department of Clinical Laboratory,the Second People's Hospital of Banan District,Chongqing 400054,China
    2. Technique Center,Ningbo Entry-Exit Inspection and Quarantine Bureau,Ningbo 315000,Zhejiang,China
    3. Reference Laboratory,Medical System Biotechnology Co.,Ltd.,Ningbo 315104,Zhejiang,China
  • Received:2018-03-20 Online:2018-08-10 Published:2018-09-07

摘要:

目的 以同位素稀释液相色谱串联质谱(ID-LC/MS/MS)为比对方法,评价酶循环法测定血清同型半胱氨酸(Hcy)的正确度。方法 采用ID-LC/MS/MS和酶循环法同时检测20份新鲜单人份血清中的Hcy浓度,对测定结果进行方法学比对和评价。采用ID-LC/MS/MS和酶循环法同时测定5个水平的工作校准品,根据比对结果调整酶循环法校准品值,直至酶循环法的测定结果与比对方法的测定结果一致。结果 采用ID-LC/MS/MS检测SRM 1950标准物质,测定结果与证书标示值一致,相对偏移<1.0%。酶循环法与ID-LC/MS/MS的线性拟合方程为Y=1.204 1X+0.364 6(r2=0.987 9),平均偏移为24.0%。调整后的酶循环法与ID-LC/MS/MS的线性拟合方程为Y=0.988 6X+0.260 3(r2=0.999 7),平均偏移为-0.09%。结论 ID-LC/MS/MS能准确测定血清Hcy水平,或许可作为Hcy常规检测系统正确度评价的比对方法。

关键词: 同型半胱氨酸, 同位素稀释液相色谱串联质谱, 酶循环法, 正确度评价

Abstract:

Objective To evaluate the accuracy of enzymatic cycling method for serum homocysteine(Hcy) based on isotope-dilution liquid chromatography-tandem mass spectrometry(ID-LC/MS/MS) as a contrast method.Methods A total of 20 fresh serum samples were determined by enzymatic cycling method and ID-LC/MS/MS,and the results were compared. The levels of serum Hcy were adjusted until the results were consistent,according to the results of 5 working calibrations,which were determined by enzymatic cycling method and ID-LC/MS/MS.Results The results of standard reference material(SRM) 1950 determined by ID-LC/MS/MS were consistent with certificate value,and the relative bias was <1.0%. The linear equation of ID-LC/MS/MS and enzymatic cycling method was Y=1.204 1X+0.364 6(r2=0.987 9),and the average bias was 24.0%. After adjusting,the linear equation was Y=0.988 6X+0.260 3(r2=0.999 7),and the average bias was -0.09%.Conclusions The levels of serum Hcy determined by ID-LC/MS/MS are accurate,which can be used for the accuracy evaluation of routine measurement procedure for serum Hcy.

Key words: Homocysteine, Isotope-dilution liquid chromatography-tandem mass spectrometry, Enzymatic cycling method, Accuracy evaluation

中图分类号: