7种半胱氨酸蛋白酶抑制剂C检测系统的分析性能评价

安崇文,李海霞   

  1. 北京大学第一医院检验科,北京 100034
  • 收稿日期:2013-04-09 出版日期:2013-12-01 发布日期:2013-12-20
  • 通讯作者: 李海霞,联系电话:010-66551785。
  • 作者简介:安崇文,男,1982年生,学士,主管技师,主要从事临床生化检验工作。

Performance evaluation of seven cystatin C analysis systems

AN Chongwen,LI Haixia.   

  1. Department of Clinical Laboratory, Peking University First Hospital, Beijing 100034, China
  • Received:2013-04-09 Online:2013-12-01 Published:2013-12-20

摘要:

目的 评估7种半胱氨酸蛋白酶抑制剂C(CysC)检测系统的分析性能。方法 应用美国临床实验室标准化协会(CLSI)系列文件评估5个颗粒增强免疫透射比浊(PETIA)检测系统和1个均相溶胶颗粒免疫法(SPIA)检测系统的精密度、正确度、空白限(LoB)、抗干扰性以及与SIEMENSBNⅡ颗粒增强免疫散射比浊(PENIA)检测系统(简称BNⅡ系统)的相关性和偏差。结果 6个检测系统(以A~F表示)和BNⅡ系统在CysC浓度为0.8~5.0mg/L时的总批内变异系数(CV)均<3.75%,总批间CV除E系统外均<5%;正确度验证显示A~E系统和BNⅡ系统测定ERM-DA471的绝对偏倚分别为0.00、0.00、1.01、-0.01、-0.50和-0.35mg/L,相对偏倚分别为0%、0%、18.43%、-0.18%、-9.12%和-6.39%;测定CAP室间质评物显示仅D系统和BNⅡ系统较为理想。A~F系统及BNⅡ系统的LoB分别为0.00、0.00、0.31、0.01、0.10、0.06和<0.05mg/L。干扰分析显示Hb≤13g/L、TG≤28mmol/L时,对A、B系统及BNⅡ系统无明显影响(干扰<±10%),其他系统均有不同程度干扰。相关分析显示A~F系统与BNⅡ系统相关性较好[相关系数(r)均>0.975,P<0.01]。偏差分析显示,A~F系统与BNⅡ系统的平均偏差分别为-0.02、-0.10、-0.31、0.05、0.04、0.36mg/L。结论 CysC各检测系统测定有证参考物质ERM-DA471的最大偏倚可达到1mg/L,最大相对偏倚为18.43%;其中PENIA在测定ERM-DA471和CAP室间质评物时结果低于PETIA;部分检测系统分析性能存在不足。

关键词: 半胱氨酸蛋白酶抑制剂C, 颗粒增强免疫透射比浊检测系统, 颗粒增强免疫散射比浊检测系统, 均相溶胶颗粒免疫检测系统

Abstract:

Objective  To evaluate the analysis performance for 7 cystatin C (Cys C ) analysis systems. Methods  A total of 5 particle-enhanced turbidimetric immunoassay (PETIA) systems and 1 sol particle immunoassay (SPIA) system were evaluated according to the documents of the Clinical and Laboratory Standards Institute (CLSI),and their correlation with particle-enhanced nephelometric immunoassay (PENIA) system from the SIEMENS BNⅡ system (BNⅡ system), precision,accuracy,limit of blank (LoB) and anti-interference ability were evaluated. Results  For 6 systems (A-F) and BNⅡ system, the total within-run coefficients of variation (CV) were all<3.75%, and except E, the total between-run CV were all<5%, when Cys C concentration was 0.8-5.0 mg/L. For A-E and BNⅡ systems,accuracy verification showed that the absolute biases for measuring ERM-DA471 were 0.00,0.00,1.01,-0.01,-0.50 and -0.35 mg/L respectively,and the relative biases were 0%,0%,18.43%,-0.18%,-9.12% and -6.39% respectively. The measurement of CAP external quality assessment materials showed that D and BNⅡ systems had desirable results. The LoB were 0.00,0.00,0.31,0.01,0.10,0.06 and<0.05 mg/L for A-F and BNⅡ systems, respectively. When Hb≤13 g/L and TG ≤28 mmol/L,A,B and BNⅡ systems were not significantly influenced (interference<±10% ). Other systems were influenced with different levels. The A-F systems correlated well with BNⅡ system, and the correlation coefficients (r) were all>0.975 (P< 0.01). Compared with BN Ⅱ system,the average deviations were -0.02,-0.10,-0.31,0.05,0.04 and 0.36 mg/L for A-F systems, respectively. Conclusions  The maximum bias can reach 1 mg/L when measuring the reference materials for Cys C analysis systems,and the maximum relative bias is 18.43%. The results of PENIA system are lower than those of PETIA system when measuring ERM-DA471 and CAP external quality assessment materials. In a part of analysis systems, analysis performance is not good.

Key words: Cystatin C, Particle-enhanced turbidimetric immunoassay system, Particle-enhanced nephelometric immunoassay system, Sol particle immunoassay system