Laboratory Medicine ›› 2018, Vol. 33 ›› Issue (11): 1018-1025.DOI: 10.3969/j.issn.1673-8640.2018.011.011

• Orginal Article • Previous Articles     Next Articles

Establishment and performance evaluation of candidate reference measurement procedure for serum total homocysteine by liquid chromatography-tandem mass spectrometry

SHEN Min1, YANG Xiaodong1, WANG Lin1, ZOU Jihua1, ZHANG Man2, ZOU Bingde1   

  1. 1. Reference Laboratory,Medical System Biotechnology Co.,Ltd.,Ningbo 315104,Zhejiang,China
    2. Department of Clinical Laboratory,Beijing Shijitan Hospital,Capital Medical University,Beijing 100038,China
  • Received:2018-01-04 Online:2018-11-30 Published:2018-11-28

Abstract:

Objective To establish a candidate reference measurement procedure for serum total homocysteine (Hcy) by liquid chromatography-tandem mass spectrometry (LC-MS/MS),and to evaluate its analytical performance. Methods Samples were prepared by a simple protein precipitation method,and LC-MS/MS was used to determine serum total Hcy quantitatively. According to the Clinical and Laboratory Standards Institute(CLSI) C62-A and C50-A documents,the performance,including linearity,limit of detection,limit of quantitation,relative matrix effect,precision and trueness,was evaluated. Results The linear range of total Hcy by LC-MS/MS was 0.5-200.0 μmol/L. The limit of quantitation and limit of detection were 0.31 nmol/g and 0.06 nmol/g,respectively. The relative matrix effects were 1.94%,1.91% and 1.78% for sera and solution mixtures in 3 different ratios(1∶1,80∶20 and 20∶80). The within-run and between-run coefficients of variation(CV)were <2% and <1%. The average recovery rates of 3 levels (30.58,49.21 and 65.42 nmol/g) were 99.8%,100.2%,100.8%, respectively. The bias of NIST standard reference material (SRM) 1950 was <1%. After the samples were processed,they were placed at room temperature [(23±2)℃] and autosampler(10℃) for 24 h, and the results were stable. Conclusions The candidate reference measurement procedure for serum total Hcy by LC-MS/MS has been established,and it has good accuracy and precision,which can be used for measurement traceability.

Key words: Total homocysteine, Isotope-dilution liquid chromatography-tandem mass spectrometry, Reference measurement procedure, Performance evaluation