Laboratory Medicine ›› 2016, Vol. 31 ›› Issue (2): 123-127.DOI: 10.3969/j.issn.1673-8640.2016.02.011

• Orginal Article • Previous Articles     Next Articles

Comprehensive evaluation of Capillarys 2 Flex Piercing glycated hemoglobin A1c determination system

XU Anping1, JI Ling1, CHEN Weidong1, XIA Yong1, HUANG Jingzhi2, YU Jing1, ZHOU Yu1, LI Lu1   

  1. 1. Shenzhen Hospital,Peking University,Shenzhen 518036,Guangdong,China
    2. Department of Laboratory Medicine,Southern Medical University,Guangzhou 510515,Guangdong,China
  • Received:2015-03-19 Online:2016-02-20 Published:2016-04-07

Abstract:

Objective To comprehensively evaluate Capillarys 2 Flex Piercing glycated hemoglobin A1c (HbA1c) determination system based on capillary electrophoresis. Methods The correlation,precision and linear range of Capillarys 2 Flex Piercing HbA1c determination system were verified. Samples with different levels of bilirubin (Bil),triglyceride (TG),hemoglobin (Hb),labile glycated hemoglobin A1c (LA1c), fetal hemoglobin(HbF) and 5 common hemoglobin variants (Hb E,Hb J-Bangkok,Hb New York,Hb G-Taipei and Hb G-Coushatta) were collected and used to evaluate the anti-interference ability of the system. The glycated hemoglobin A2 (HbA2) values were determined by Capillarys 2 Flex Piercing HbA1c determination system,which was used to evaluate the correlation with capillary electrophoresis. The 40 samples whose HbA1c values were assigned by the National Glycohemoglobin Standardization Program (NGSP) were determined by Capillarys 2 Flex Piercing HbA1cdetermination system,and the result difference between Capillarys 2 Flex Piercing HbA1c determination system and NGSP reference laboratory results was analyzed,in order to verify accuracy. Results Capillarys 2 Flex Piercing HbA1c determination system and Variant Ⅱ Turbo 2.0 HbA1c analyzer were correlated well (r =0.997,P=0.000). The total precision[coefficient of variation (CV)] was <2.00%. The linearity proved to be good within a range of HbA1c values of 3.8%-17.1%. Capillarys 2 Flex Piercing HbA1c determination system was not influenced at Bil≤355 μmol/L,TG≤25.3 mmol/L,Hb 28-182 g/L,LA1c 1.2%-7.8% and HbF≤22.3%. The 5 common hemoglobin variants (Hb E,Hb J-Bangkok,Hb New York,Hb G-Taipei and Hb G-Coushatta) could be separated. HbA2 values from Capillarys 2 Flex PiercingHbA1c determination system and capillary electrophoresis were correlated well (r =0.993,P=0.000). The result differences with NGSP reference laboratory results were all <6%. Conclusions The analysis performance of Capillarys 2 Flex Piercing HbA1c determination system meets the requirements of routine clinical application,and the determination system also provides practical HbA2 values for thalassemia screening. Capillarys 2 Flex Piercing HbA1c determination system is a multifunctional HbA1c determination system.

Key words: Glycated hemoglobin A1c, Capillary electrophoresis, Capillarys 2 Flex Piercing glycated hemoglobin A1c determination system, The National Glycohemoglobin Standardization Program

CLC Number: