Laboratory Medicine ›› 2015, Vol. 30 ›› Issue (6): 603-606.DOI: 10.3969/j.issn.1673-8640.2015.06.015

Previous Articles     Next Articles

Performance evaluation of Cobio Palio 300 urine biochemical analyzer for the determinations of 5 urinary micro-proteins

HUANG Yan, XU Xiaorong, ZHANG Jue   

  1. Department of Clinical Laboratory, Shuguang Hospital, Shanghai 201203, China
  • Received:2014-12-15 Online:2015-06-30 Published:2015-07-03

Abstract: Objective

To evaluate the performance of Cobio Palio 300 urine automatic biochemical analyzer (Palio 300) for the determinations of 5 kinds of urinary micro-proteins.

Methods

By Palio 300 and Siemens BN-prospec, 120 urine samples were determined for urinary micro-albumin (mAlb), α1 microglobulin (α1-MG), β2 microglobulin (β2-MG), transferrin (TRF) and immunoglobulin G (IgG). The results were compared and analyzed statistically. In addition, the medium-value and high-value mixed urine was used for these items. Each sample was determined repeatedly for 20 times, and the coefficient of variation (CV) was calculated.

Results

Four items showed CV<6.00%, except β2-MG CV=8.66%. The results were consistent with those of Siemens BN-prospec. The correlation coefficients (r) of 5 items were >0.9 (mAlb 0.953, α1-MG 0.915, β2-MG 0.927, TRF 0.918 and IgG 0.935, P<0.001).

Conclusions

Palio 300 has good repeatability, and its result is consistent with Siemens BN-prospec in the determinations of 5 urinary micro-proteins.

Key words: Micro-albumin, α1 microglobulin, β2 microglobulin, Transferrin, Immunoglobulin G, Urine, Cobio Palio 300 urine biochemical analyzer, Turbidimetric immunoassay

CLC Number: