Laboratory Medicine ›› 2019, Vol. 34 ›› Issue (5): 443-446.DOI: 10.3969/j.issn.1673-8640.2019.05.013

Previous Articles     Next Articles

IFCC Secondary Reference Measurement Procedure for 4 HbA1c determination systems

LIU Wenbin, JU Yi(), TANG Liping, WANG Meijuan, OU Yuanzhu, YU Xiaoxuan, LI Qing   

  1. Shanghai Center for Clinical Laboratory,Shanghai 200126,China
  • Received:2018-06-22 Online:2019-05-30 Published:2019-05-28

Abstract:

Objective To improve the traceability of glycated hemoglobin A1c(HbA1c) in Shanghai Center for Clinical Laboratory with establishing the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) Secondary Reference Measurement Procedure,through tracing from the IFCC Primary Reference System. Methods Using 4 HbA1c determination systems,including HA-8180,Hb-9210,Capillarys 2 Flex Piercing (Cap2) and MQ2000-PT with their original reagents and calibrators,the results of 24 samples sent by IFCC were compared with the target values assigned by the IFCC Primary Reference System. The performance of the 4 determination systems were evaluated by the IFCC rules. Results The coefficients of variation (CV) of the 4 determination systems were 1.1% for HA-8180,2.0% for Hb-9210,2.2% for Cap2 and 2.0% for MQ2000-PT. The imprecisions were 0.6,1.0,1.1 and 1.0 mmol/mol,respectively. The |biases| at the level of 50 mmol/mol were 1.0,1.1,0.4 and 1.0 mmol/mol,respectively. The total errors (TE) of whole year were 2.2,3.1,2.6 and 3.2 mmol/mol,respectively. According to the rules,HA-8180 got a "Sliver" score,and Hb-9210,Cap2 and MQ2000-PT got "Bronze" score. Conclusions The 4 determination systems demonstrate traceability to the IFCC Secondary Reference Measurement Procedure.

Key words: Glycated hemoglobin A1c, Reference measurement procedure, Total error, Biological variation

CLC Number: