›› 2012, Vol. 27 ›› Issue (3): 178-180.
Previous Articles Next Articles
Received:
Online:
Published:
Abstract: Objective To evaluate and improve the method for detecting urine and cerebrospinal fluid proteins by sulfosalicylic acid turbidimetry using automatic biochemical analyzer. Methods The improved sulfosalicylic acid turbidimetry using automatic biochemical analyzer(I-SS assay) and its parameters were established. The precision,linear range and recovery rate were evaluated and compared with the traditional sulfosalicylic acid turbidimetry(T-SS assay). Results The withinrun and betweenrum coefficients of variation(CV) were 1.2%-1.5% and 2.1%-3.5%. The linear range was 0.037-2.38 g/L. The recovery rate was 85.6%-109.2%. The correlation with the T-SS assay was Y I-SS = 0.980 4X T-SS +0.010 8, and the correlation coefficient (r) was 0.982. Conclusions The I-SS assay is sensitive and accurate, and it is suitable to automatic biochemical analyzer.
Key words: Urine protein, Sulfosalicylic acid, Turbidimetry, Automatic biochemical analyzer
XIAO Chunhai;HUANG Minjie;WU Juanfang;ZHAO Yifan;SHEN Yuping. Developement a method for urine and cerebro-spinal fluid proteins by salicylsulfonic acid teturbidimetry in automatic biochemical analyzer[J]. , 2012, 27(3): 178-180.
0 / / Recommend
Add to citation manager EndNote|Ris|BibTeX
URL: https://www.shjyyx.com/EN/
https://www.shjyyx.com/EN/Y2012/V27/I3/178