›› 2012, Vol. 27 ›› Issue (3): 178-180.

Previous Articles     Next Articles

Developement a method for urine and cerebro-spinal fluid proteins by salicylsulfonic acid teturbidimetry in automatic biochemical analyzer

  

  1. Department of Clinical Laboratory, Jinshan Branch of Shanghai Sixth People′s Hospital,Shanghai 201500,China
  • Received:2012-01-12 Online:2012-03-30 Published:2012-03-16

Abstract: Objective To evaluate and improve the method for detecting urine and cerebrospinal fluid proteins by sulfosalicylic acid turbidimetry using automatic biochemical analyzer. Methods The improved sulfosalicylic acid turbidimetry using automatic biochemical analyzer(I-SS assay) and its parameters were established. The precision,linear range and recovery rate were evaluated and compared with the traditional sulfosalicylic acid turbidimetry(T-SS assay). Results The withinrun and betweenrum coefficients of variation(CV) were 1.2%-1.5% and 2.1%-3.5%. The linear range was 0.037-2.38 g/L. The recovery rate was 85.6%-109.2%. The correlation with the T-SS assay was Y I-SS = 0.980 4X T-SS +0.010 8, and the correlation coefficient (r) was 0.982. Conclusions The I-SS assay is sensitive and accurate, and it is suitable to automatic biochemical analyzer.

Key words: Urine protein, Sulfosalicylic acid, Turbidimetry, Automatic biochemical analyzer