Performance evaluation of galectin-3 detection by chemiluminescence and its preliminary application in heart failure

ZHENG Xiaorong1, WAN Jun2, ZHANG Liangying3, LI Yongshu4, HAN Xuejing4, SHI Ping4, JIA Kegang4   

  1. [1. Tianjin Medical University Cardiovascular College,Tianjin 300457,China
    2. The First People's Hospital of Urumqi Economic and Technological Development Area(Toutunhe Zone),Urumqi 830023,Xinjiang,China
    3.College of Laboratory Medicine,Hebei North University,Zhangjiakou 075000,Hebei,China
    4. Tianjin Medical University Cardiovascular College TEDA International Cardiovascular Hospital,Tianjin 300457,China]
  • Received:2018-07-26 Online:2019-04-30 Published:2019-05-17

Abstract:

Objective To evaluate the performance of galectin-3(Gal-3) detection by chemiluminescence and its preliminary application in heart failure(HF). Methods According to related documents,the performance of Gal-3 detection by chemiluminescence [limit of blank(LoB),limit of detection(LoD),limit of quantitation(LoQ),precision,linear range and interference] was evaluated,and the manufacturer's reference interval was verified. The effects of age,sex and specimen type on Gal-3 detection were compared. A total of 55 patients with HF(HF group) and 80 healthy subjects(healthy control group) were enrolled,and their plasma samples were collected for Gal-3 detection. Receiver operating characteristic(ROC) curve was used to analyze the role of Gal-3 in HF diagnosis. Results The Gal-3 kit(chemiluminescence) got a satisfying result in verification on LoB,LoD and LoQ,with the values of 0.80,1.0 and 1.0 ng/mL by the manufacturer,respectively. The repeat precision expressed by coefficient of variation(CV) was 1.9%-4.8%,and the within-laboratory precision expressed by CV was 2.6%-6.1%,and the manufacturer's precision got verification. In the range of 4.07-113.11 ng/mL,there was a good linearity of Gal-3 detection by chemiluminescence. Serum Gal-3 level determined by chemiluminescence would not be affected in samples with hemoglobin level <20 g/L,total bilirubin(TB)<403.2 μmol/L,direct bilirubin(DBil) < 214.08 μmol/L or the turbidity of chyle <7 450 FTU. The plasma Gal-3 level was higher than that in serum (P<0.05). There was no statistical significance in Gal-3 levels between males and females(P>0.05). The research subjects were classified by sample types(serum and plasma) and ages. In serum group,Gal-3 level had statistical significance in 3 age subgroups(20~<45-year-old group,45~<60-year-old group and ≥60-year-old group)(P<0.05),and there was statistical significance for plasma Gal-3 levels in 20~<60-year-old group and ≥60-year-old group(P<0.05). Gal-3 levels increased with ages. The level of Gal-3 was higher in patients with HF than that in healthy control group(P<0.001). The area under ROC curve (AUC) of Gal-3 for HF diagnosis was 0.969,and the optimal cut-off value was 20.04 ng/mL,with a sensitivity of 85.5% and a specificity of 100.0%. Conclusions The analytical performance of Gal-3 detection by chemiluminescence is good. Age has a certain effect on Gal-3 levels. Gal-3 plays a role in the auxiliary diagnosis of heart failure.

Key words: Galectin-3, Chemiluminescence, Analytical performance, Heart failure