Determination of high-concentration enzyme samples by FLEX mode-TOSHIBA TBA-120FR automatic biochemical analyzer

TIAN Honglun, YIN Ke, HE Jun, LEI Wenmei, ZHANG Ziping   

  1. Department of Clinical Laboratory,Guizhou Second Provincial People's Hospital,Guiyang 550004,Guizhou,China
  • Online:2016-08-31 Published:2016-09-01

Abstract:

Objective To investigate the performance of FLEX mode-TOSHIBA TBA-120FR automatic biochemical analyzer in the determination of high-concentration enzyme samples. Methods The high-concentration alanine aminotransferase (ALT)and aspartate aminotransferase(AST)samples(ALT:2 185-5 240 U/L,AST:2 520-6 625 U/L)were determined by FLEX mode and manual dilution mode,the relative deviation of 2 modes was calculated,and the 2 modes were analyzed by linear regression analysis. Results The relative deviation was <6% between FLEX mode and manual dilution mode,which met the health industry standard WS/T403-2012 requirements. The linear regression analysis showed that the 2 modes had a good correlation [ALT:coefficient of determination(R2)=0.999,P<0.001;AST:R2=0.999,P<0.001]. The medical decision levels of high-concentration ALT and AST generated into regression equation,and the relative deviations were 1.9% and 2.2% for calculating medical decision levels. Less than half of allowable total error from the Clinical Laboratory Improvement Amendments in 1988(CLIA'88)was used as clinical acceptable level. Conclusions The reportable range of enzyme activity determination is expanded by FLEX mode,which can ensure obtaining results accurately and rapidly and improve work efficiency, and FLEX mode is of significance.

Key words: FLEX mode, Alanine aminotransferase, Aspartate aminotransferase

CLC Number: