Evaluation on the performance of Reflotron-Hemoglobin dry-chemistry system

CHEN Liting, WANG Jianbiao, LIN Xiaoyi, SHI Hourong.   

  1. Department of Clinical Laboratory, Ruijin Hospital, Shanghai Jiaotong University School of Medicine, Shanghai 200025, China
  • Received:2014-01-10 Online:2014-05-30 Published:2014-05-27

Abstract: Objective To evaluate Reflotron-Hemoglobin dry-chemistry system for the measurement of hemoglobin. Methods The within-run and between-run precisions, linearity, correlation and reference range of Reflotron- Hemoglobin dry-chemistry system was verified. Results The within-run precision was 2.54%-2.84%, and the between-run precision was 0.68%-3.38% for Reflotron-Hemoglobin dry-chemistry system, respectively. The linearity range was 52-191 g/L . The results of Reflotron-Hemoglobin dry-chemistry system were correlated well with those of Beckman-Coulter LH750 system, and the correlation equation wan Y=1.031 6X-2.796 9. The verification coincidence rate for reference range was 100%. Conclusions The within-run and between-run precisions of Reflotron-Hemoglobin dry-chemistry system for the determination of hemoglobin are good with accurate results for both normal and abnormal samples. The reference range can be used in this system.

Key words: Hemoglobin, Performance evaluation, Precision, Linearity