检验医学 ›› 2020, Vol. 35 ›› Issue (5): 464-469.DOI: 10.3969/j.issn.1673-8640.2020.05.016

• 技术研究与评价·论著 • 上一篇    下一篇

均相法检测脂蛋白残粒胆固醇的建立及初步临床应用

王蓉1, 黄克亮1, 张建华2, 张欣1   

  1. 1. 上海交通大学医学院附属第九人民医院黄浦分院检验科,上海 200011
    2. 上海交通大学医学院化学教研室,上海 200025
  • 收稿日期:2019-11-28 出版日期:2020-05-30 发布日期:2020-06-17
  • 作者简介:null

    作者简介:王 蓉,女,1968年生,硕士,副主任技师,主要从事医学检验和实验室管理工作。

  • 基金资助:
    上海市黄浦区卫生健康委员会科研项目(HKM201815)

Establishment and preliminary clinical application of homogeneous method for the determination of remnant lipoprotein-cholesterol

WANG Rong1, HUANG Keliang1, ZHANG Jianhua2, ZHANG Xin1   

  1. 1. Department of Clinical Laboratory,Huangpu Branch,the Ninth People's Hospital Affiliated to Shanghai Jiao Tong University School of Medicine,Shanghai 200011,China
    2. Department of Chemistry,Shanghai Jiao Tong University School of Medicine,Shanghai 200025,China
  • Received:2019-11-28 Online:2020-05-30 Published:2020-06-17

摘要:

目的 建立均相法检测脂蛋白残粒胆固醇(RLP-C)的方法,并进行初步评估。方法 以聚氧乙烯-聚氧丁烯嵌段共聚物(POE-POB)为屏蔽剂,特异性屏蔽非脂蛋白残粒,以此为基础,优化实验条件,建立检测RLP-C的均相法。将自建方法与参比方法[日本协和公司RLP-C试剂(均相法)]进行比对,采用自建方法测定116例混合性高脂血症患者和130名正常对照者的血清RLP-C浓度,同时测定血清三酰甘油(TG)、总胆固醇(TC)、载脂蛋白C2(apo C2)浓度。采用Pearson相关分析评估各项目之间的相关性。结果 自建方法的最佳缓冲体系为3-(N-玛代琳)丙磺酸(MOPS)缓冲体系,最适pH值为6.8,POE-POB最适浓度为8 g/L,检测延迟时间为5 min,测定时间为5 min。自建方法的批内平均变异系数(CV)为3.37%,批间平均CV为4.91%,试剂在4 ℃条件下可稳定1个月。参比方法与自建方法的检测结果稍有差异,但2种方法的相关性良好(r=0.980),Bland-Altman散点图显示,98例样本中仅有4例(4.08%)不在±1.96s范围内。混合性高脂血症组血清RLP-C、TG、apo C2、TC浓度均高于正常对照组(P<0.001)。RLP-C与TG呈正相关(r=0.496,P<0.001),与apo C2、TC均无相关性(r值分别为0.157、-0.090,P>0.05)。结论 自建方法简便、快捷,可用于全自动生化分析仪,适用于临床常规检测。

关键词: 脂蛋白残粒胆固醇, 均相法, 聚氧乙烯-聚氧丁烯嵌段共聚物, 高脂血症

Abstract:

Objective To establish homogeneous method for determining remnant lipoprotein-cholesterol(RLP-C),and to evaluate it primarily. Methods A homogeneous method for the determination of RLP-C was established based on the screening of non-lipoproteins with polyoxyethylene-polyethylene oxide butene(POE-POB) as screening agent. The self-established method was compared with the reference method [RLP-C reagent(homogeneous method) of Japan Concorde company]. Serum RLP-C and triglyceride(TG),total cholesterol(TC),apolipoprotein C2(apo C2) were determined in 116 patients with mixed hyperlipidemia and 130 healthy controls. Pearson correlation analysis was used to evaluate their correlations. Results The optimal buffer system was 3-morpholinopropanesulfonic acid(MOPS) buffer system. The optimal pH value was 6.8,the optimal concentration of POE-POB was 8 g/L,and the determination delay time was 5 min. Intra-self established method coefficient of value(CV) was 3.37%,and inter-CV was 4.91%. The reagent was stable for 1 month at 4 ℃. The results of reference method and self-established method were slightly different,but the correlation between the 2 methods was good(r=0.980). Bland-Altman scatter plot showed that 4(4.08%) samples were not within ±1.96s. The concentrations of RLP-C,TG,apo C2 and TC in mixed hyperlipidemia group were higher than those in healthy control group(P<0.001). RLP-C was positively correlated with TG(r=0.496,P<0.001),but it was not correlated with apo C2 and TC(r=0.157 and -0.090,P>0.05). Conclusions The self-established method is simple and rapid,which can be used in automatic biochemical analyzer and be used as clinical routine test.

Key words: Remnant lipoprotein-cholesterol, Homogeneous method, Polyoxyethylene-polyethylene oxide butene, Hyperlipidemia

中图分类号: