Investigation on the cut-off values of tandem mass spectrometry for neonatal inherited metabolic diseases in China

HE Falin1, LIN Bo2, WANG Wei1, ZHONG Kun1, YUAN Shuai1, DU Yuxuan1, WANG Zhiguo1   

  1. 1. National Center for Clinical Laboratories,Beijing Engineering Research Center of Laboratory Medicine,Beijing Hospital,National Center of Gerontology,Beijing 100730,China
    2. College of Computer Science and Technology,Zhejiang University,Hangzhou 310013,Zhejiang,China
  • Received:2019-01-10 Online:2020-05-30 Published:2020-06-17

Abstract:

Objective To investigate the cut-off values of 16 indicators for screening neonatal inherited metabolic diseases by tandem mass spectrometry in China,and to provide a reference on setting cut-off values for screening neonatal inherited metabolic diseases. Methods The data related to the cut-off values of 4 amino acids and 12 acylcarnitines from 141 neonatal screening laboratories,including laboratory coding,cut-off value,source of cut-off value,methodological principle,instrument,reagent,calibrator,modifying or not,were collected. The effects of different methods and sources of cut-off values on the 16 cut-off values were analyzed. Results A total of 114 laboratories submitted their data. Totally,70.18% laboratories' cut-off values were from reagent manufacturer's instructions,about 16% laboratories' cut-off values were determined by own laboratory,and about 14% laboratories' cut-off values were from literature reports. The proportion of laboratories with modified cut-off values was 49.12%-53.51%. The cut-off values had statistical significance between derivative method and non-derivative method(P<0.05). The coefficient of variation of derivative method was distributed in 25.26%-172.05%,and the dispersion degree of unmodified tangent value was greater than that of modified tangent value. The coefficient of variation of non-derivative method was distributed in 18.15%-200.92%,and the unmodified tangent values were consistent with the dispersion degree of modified tangent values except low citrulline values. Conclusions There are differences in the cut-off value of tandem mass spectrometry in screening neonatal inherited metabolic diseases in China,and the cut-off value should be evaluated and validated regularly.

Key words: Neonatal inherited metabolic disease, Cut-off value, Amino acid, Acylcarnitine