Laboratory Medicine ›› 2020, Vol. 35 ›› Issue (2): 137-141.DOI: 10.3969/j.issn.1673-8640.2020.02.010

Previous Articles     Next Articles

Evaluation on chemiluminescence microparticle immunoassay in the detection of anti-TP antibody

TONG Minghong1, WU Xiaohui1, WANG Zhenhua2, ZHANG Tingying1, WEI Xin1, CHEN Cuiwen1, LI Ying1, CUI Xiangming1, SHENG Huiming1(), HUANG Qiang3()   

  1. 1. Tongren Hospital,Shanghai Jiao Tong University School of Medicine,Shanghai 200336,China
    2. Shanghai Jiao Tong University School of Medicine,Shanghai 200025,China
    3. Community Health Service Center of Shanghai Putuo District Changshou Street,Shanghai 200063,China
  • Received:2019-05-08 Online:2020-02-29 Published:2020-03-22

Abstract:

Objective To investigate the performance and signal-to-cutoff(S/CO)ratio setting scheme for the detection of anti-Treponema pallidum(TP) antibody by chemiluminescence microparticle immunoassay (CMIA). Methods Anti-TP antibody was detected in 116 patients by CMIA,which was confirmed by fluorescent treponemal antibody absorption test(FTA-ABS). The results of FTA-ABS were used as gold standard. The performance of CMIA was evaluated by receiver operating characteristic(ROC) curve. Results There were 116 patients detected by CMIA,including 98 patients with S/CO>1.1 and 18 patients with S/CO≤1.1,of which 70 cases were positive,16 cases were weakly positive,and 30 cases were negative by FTA-ABS. The clinical diagnosis results were positive in 58 cases and negative in 58 cases. According to the FTA-ABS results,the area under curve of anti-TP antibody detected by CMIA was 0.885,and the Youden index was 0.757. The optimal cut-off value of anti-TP antibody S/CO ratio of CMIA was 3.66,the sensitivity was 95.7%,and the specificity was 80.0%. When the positive predictive value of anti-TP antibody detected by CMIA was ≥95%,the S/CO ratio was 4.25. According to the clinical diagnosis results,the area under curve of anti-TP antibody detected by CMIA was 0.944,and the Youden index was 0.764. The optimal cut-off value of anti-TP antibody S/CO ratio of CMIA was 4.25,the sensitivity was 87.9%,and the specificity was 88.5%. Conclusions CMIA has high sensitivity,no missed diagnosis and certain case misdiagnosis rate,so it is suitable for preliminary screening.

Key words: Anti-Treponema pallidum antibody, Chemiluminescence microparticle immunoassay, Treponema pallidum particle agglutination assay, Fluorescent treponemal antibody absorption test, Syphilis