Laboratory Medicine ›› 2017, Vol. 32 ›› Issue (11): 1046-1050.DOI: 10.3969/j.issn.1673-8640.2017.011.024

• Orginal Article • Previous Articles     Next Articles

Establishment of a liquid chromatography-tandem mass spectrometry for the determination of plasma 3-methoxytyramine

QIN Jiaqian, PENG Yingfei, CHEN Fangjun, SHAO Wenqi, WU Jiong, ZHANG Chunyan, WANG Beili, GUO Wei, PAN Baishen   

  1. Department of Clinical Laboratory,Zhongshan Hospital,Fudan University,Shanghai 200032,China
  • Received:2017-02-13 Online:2017-11-30 Published:2017-12-07

Abstract:

Objective To establish a liquid chromatography-tandem mass spectrometry(LC-MS/MS) for plasma 3-methoxytyramine(3-MT) determination. Methods The separation condition of LC-MS/MS(chromatographic column,column temperature and pH value)was improved. The linear range,recovery rate,precision,lower limit of quantitation and stability of LC-MS/MS were evaluated. Results 3-MT was separated by BEH HILIC chromatographic column using 3-MT-d4 as internal standard. The mobile phase consisted of 100 mmol/L ammonium formate aqueous solution(pH value was 3)and acetonitrile. The column temperature was 35 ℃. The linear range of 3-MT determination were 5-1 000 pg/mL. The limit of quantitation was 5 pg/mL. The inter-day and between-run coefficients of variation(CV) were <6% and <7%. The recovery rate was 97.1%-109.3%. As 3-MT was unstable in room temperature,it should be transported under 4 ℃. Conclusions A reliable LC-MS/MS for the determination of 3-MT has been established,with sensitivity and accuracy.

Key words: 3-Methoxytyramine, Liquid chromotography-tandem mass spectrometry, Pheochromocytoma

CLC Number: