UF-1000i全自动尿液分析仪的细菌通道检测与中段尿培养的符合率分析

冯玉青, 黄素素, 罗靖, 顾国龙   

  1. 广西中医药大学附属瑞康医院检验科,广西 南宁 530011
  • 收稿日期:2017-06-03 发布日期:2021-03-18
  • 作者简介:null

    作者简介:冯玉青,女,1978年生,硕士,主管技师,主要从事临床微生物学检验工作。

Consistency rate of UF-1000i automatic urine analyzer bacterial channel and mid-stream urine culturing

FENG Yuqing, HUANG Susu, LUO Jing, GU Guolong   

  1. Department of Clinical Laboratory,Ruikang Hospital,Guangxi University of Chinese Medicine,Nanning 530011,Guangxi,China
  • Received:2017-06-03 Published:2021-03-18

摘要:

目的 分析UF-1000i全自动尿液分析仪(简称UF-1000i)的细菌通道检测与中段尿培养的符合率,为临床尿路感染(UTI)的早期经验性用药提供依据。方法 对中段尿样本同时进行细菌培养和尿有形成分分析,培养阳性的样本采用Vitek 2 Compact微生物鉴定仪进行菌种鉴定,尿有形成分分析采用UF-1000i的细菌通道进行检测。结果 以中段尿培养为金标准,UF-1000i的细菌通道检测革兰阳性球菌和革兰阴性杆菌的符合率、敏感性、特异性分别为90.7%、60.8%、93.5%和93.3%、76.7%、96.8%(P<0.05)。结论 UF-1000i的细菌通道检测在UTI患者早期选择抗菌药物方面有较好的参考价值,特别是对革兰阴性杆菌引起的UTI,有更高的参考价值。

关键词: UF-1000i全自动尿液分析仪的细菌通道, 中段尿培养, 符合率

Abstract:

Objective To study the consistency rate of UF-1000i automatic urine analyzer bacterial channel and mid-stream urine culturing,and to provide a reference for the early empirical use of antibiotics for urinary tract infection(UTI). Methods Urine specimens were determined for bacterial culturing and urinary sedimentation analysis. Bacterial culturing was performed by Vitek 2 Compact. Urinary sedimentation analysis was performed by UF-1000i becterial channel. Results Th. Results of mid-stream urine culturing were as gold standard. For UF-1000i bacterial channel,the consistency rate,sensitivity and specificity of Gram-positive coccus were 90.7%,60.8% and 93.5%,and those of Gram-negative bacillus were 93.3%,76.7% and 96.8% (P<0.05).Conclusions UF-1000i bacterial channel plays a role in the early clinical selection of antibiotics,especially in UTI caused by Gram-negative bacillus.

Key words: UF-1000i automatic urine analyzer bacterial channel, Mid-stream urine culturing, Consistency rate