Laboratory Medicine ›› 2019, Vol. 34 ›› Issue (2): 159-161.DOI: 10.3969/j.issn.1673-8640.2019.02.014

Previous Articles     Next Articles

Immature platelet fraction determination by Sysmex XN automated hematology analyzer for thrombocytopenia

LIN Xiaoyi, CHEN Liting, SHEN Yun   

  1. Department of Clinical Laboratory,Ruijin Hospital,Shanghai Jiaotong University School of Medicine,Shanghai 200025,China
  • Received:2018-03-29 Online:2019-02-28 Published:2019-02-28

Abstract:

Objective To evaluate the role of immature platelet fraction(IPF%) determination by Sysmex XN automated hematology analyzer for thrombocytopenia. Methods The platelet(PLT) counts and IPF% of 200 patients with thrombocytopenia [45 patients with primary immune thrombocytopenia(ITP),40 patients with aplastic anemia(AA),60 patients undergoing chemotherapy(CT) and 55 patients with chronic liver diseases(CLD)] and 50 healthy subjects(healthy control group) were determined by Sysmex XN automated hematology analyzer. The difference of IPF% was analyzed by GraphPad Prism 6 software,and receiver operating characteristic(ROC) curve analysis was utilized to demonstrate the diagnostic efficiency of IPF% in the assessment of thrombocytopenia. Results The PLT counts of ITP,AA,CT and CLD groups were lower than that of healthy control group(P<0.05). However,there was no statistical significance between ITP and AA groups and between CT and CLD groups(P>0.05). IPF% of ITP group was higher than those of healthy control,AA,CT and CLD groups(P<0.05),while there was no statistical significance between AA,CT and CLD groups and healthy control group(P>0.05). When the cut-off value of IPF% was 5.8%,the sensitivity was 82.2%,the specificity was 86.8%,and the area under curve(AUC) was 0.92. Conclusions IPF% determination by Sysmex XN automated hematology analyzer plays a role for discriminating hypo- from hyper-proliferative thrombocytopenias.

Key words: Immature platelet fraction, Thrombocytopenic disorder, Automated hematology analyzer

CLC Number: