The evaluation of enzymatic method for the determination of serum sodium ion

  

  • Received:2012-08-03 Revised:2012-07-24 Online:2012-10-30 Published:2012-10-12

Abstract: Objective To evaluate the performance of enzymatic method for serum sodium ion determination.  Methods  The accuracy,precision,off cap stability,linearity,correlation with ion selective electrode method,recovery rate and anti-interference were evaluated by the reagent of enzymatic method for sodium ion determination.  Results The coefficient of variation (CV) was<2%. Calibration period reached 5 d. Accuracy met Randox′s quality control. Linearity scope was 100-160mmol/L. Coefficient of correlation (r) was 0.989 9, and enzymatic method was correlated well with ion selective electrode method. The recovery rate was 100.9%. No interference appeared when bilirubin≤600 μmol/L,triglyceride≤20 mmol/L,vitamin C≤0.5 g/L,hemoglobin≤10 g/L,ammonium ion ≤25 mmol/L,potassium ion ≤10 mmol/L, magnesium ion ≤5 mmol/L,calcium ion≤5 mmol/L,zinc ion≤50 mmol/L and copper ion≤50 mmol/L.  Conclusions Enzymatic method meets clinical requirements with good repeatability,accuracy,stability,linearity and high anti-interference.

Key words: Sodium ion, Enzymatic method, Ion selective electrode method